9
4. Press
MI/
▲
or MR/
▼
to select: NO retains current settings
or
YES clears all calibrations and its
data.
5.
Press ENTER to confirm and return to subgroup menu.
6.
Press CAL/MEAS to return to measurement mode.
7.5
Setting the temperature coefficient (P5.1)
The temperature coefficient is the amount of change in conductivity per degree of temperature; it is
expressed in percent per °C. Entering the exact temperature coefficient of your solution lets you
accurately compensate temperature for almost any solution. You can adjust 0.0 to 10.0 % per °C.
Meter default is 2.1% per °C. By setting the temperature coefficient to 0.0%, the meter will not apply
any compensation but will display actual temperature.
1. Press
SETUP
2. Press
MI/
▲
or MR/
▼
until you view parameter P5.0.
3.
Press ENTER to select P5.1 (display shows “t.CO”).
4. Press
ENTER.
5. Press
MI/
▲
or MR/
▼
to select the temperature coefficient of your solution.
6.
Press ENTER to confirm and return to subgroup menu.
7.
Press CAL/MEAS twice to return to the measurement mode.
7.6
Setting the normalisation temperature (P5.2)
Your meter will normalize its conductivity measurements to a standard temperature that you can
select. You can adjust the normalization temperature from 15°C to 30°C (59 to 86 °F). Meter default is
25.0°C (77 °F).
1. Press
SETUP.
2. Press
MI/
▲
or MR/
▼
until you view parameter P5.0.
3.
Press ENTER three times to select P5.2 (display shows “t.nr”).
4. Press
ENTER.
5. Press
MI/
▲
or MR/
▼
to select either the normalization temperature.
6.
Press ENTER to confirm and return to subgroup menu.
7.
Press CAL/MEAS to return to measurement mode.
7.7
Selection of cell constant (P6.0)
The meter lets you select a cell constant of K = 1.0, 10, or 0.1. The cell included with your meter has a
cell constant of K=1.0.
1. Press
SETUP.
2. Press
MI/
▲
or MR/
▼
until you view parameter P6.0.
3. Press
ENTER.
4. Press
MI/
▲
or MR/
▼
to select the cell constant between K = 1.0, 0.1, or 10.
5.
Press ENTER to confirm selection and return to the subgroup menu.
6.
Press CAL/MEAS to return to measurement mode.
7.8
Setting the real-time clock (P7.0)
The meter features a real-time calendar and clock. This helps you to meet the Good Laboratory
Practice (GLP) standards.
1. Press
SETUP.
2. Press
MI/
▲
or MR/
▼
until you view parameter P7.0.
3. Press
ENTER.
4. Press
MI/
▲
or MR/
▼
to select century: “19—“ or “20—“. The century digits will flash.
5.
Press ENTER to confirm and move to “year” selection.
6. Press
MI/
▲
or MR/
▼
to select year. The “year” digits will flash.
7.
Press ENTER to confirm and move to “month” selection.
8. Press
MI/
▲
or MR/
▼
to select month. The “month” digits will flash.
9.
Press the ENTER
to confirm and move to “date” selection.
10. Press
MI/
▲
or MR/
▼
to select date. The “date” digits will flash.
11. Press ENTER to confirm and move to “hour” selection.
12. Press
MI/
▲
or MR/
▼
to select hour. Note the “AM” / “PM” indicator in the display. The “hour”
digits will flash.