9. Manual control mode

You can control devices connected to Relay A or Relay B from the front panel of this
controller. Automatic mode is the mode that the controller normally enters when
you switch it on.  In Automatic mode, the controller’s set-point values actuate the
relays. 

In Manual mode, you have manual control of your relays so you can prime your
pump or check pump status without operating the entire system.

To enter manual mode:

1.

Press the Relay Control (auto/manu) key.

The upper display will show “000”;

the lower display will show “S.Cd” to prompt you to enter the Advanced Set-up
Code.

2. 

Press the 

and 

keys

to scroll the upper display until it reads “22”. 

3. 

Press the ENTER key. 

The Manual indicator by the Relay Control key will light.

NOTE: 

Pressing ENTER at a value other than “22” will cause the controller to revert
to the Measurement mode, and the relays will remain in automatic mode.

4.  

Press the Relay Selection key

to select either Relay A or Relay B. The LED next

to the currently selected relay (A or B) will light. The upper display will read the
current measured value. The lower display will show “oFF” or “on” depending
on the relay status of the currently selected relay. 

5.

Press the 

and 

keys

to change the Relay on/off status. The LED indicators at

the right of the controller will also change between Red and Green to indicate
Relay status.

NOTE: 

If you wish to manually change the status of both relays, press the RELAY
SELECTION key at this point and repeat step 5 for the second relay. This
first relay will remain under manual control while you are setting the second
relay.

6.

Press the RELAY CONTROL (auto/manu) key

to return to Measurement mode.

The relays are now back under automatic control.

10. Technical Specifications

Range no.

Range

Resolution

Cell Constant (K)

Conductivity controllers

1

0.000 to 2.000 µS

0.001 µS

0.01

2

0.00 to 20.00 µS

0.01 µS

0.01

3

0.00 to 20.00 µS

0.01 µS

0.10

4

0.0 to 200.0 µS

0.1 µS

0.10

5

0.0 to 200.0 µS

0.1 µS

1.0

6

0 to 2000 µS

1 µS

1.0

7

0 to 5000 µS

5 µS

1.0

8

0.00 to 20.00 mS

0.01 mS

1.0

9

0.0 to 200.0 mS

0.1 mS

10.0

0

0.0 to 200.0 mS

0.1 mS

1.0

Resistivity controllers

1

0.0 to 20.00 M

0.01 M

0.01

2

0.000 to 1.999 M

0.001 M

0.10

61

Control type:

on/off

Control accuracy:

±0.5% full-scale

Number of inputs:

one

Number of set points:

two

Output: 

Control: 2 SPDT relays, 6 A @ 110 VAC,

250 VAC max

Alarm: 1 SPDT relay, 6 A @ 110 VAC, 

250 VAC max

Current: isolated 0-20/4-20 mA, with

20% to 100% full-scale scalable 
boundaries, 600 

max. load

Hysteresis (dead band): 

0 to 10% of range

Relay delay: 

selectable, 0 to 2000 seconds

Electrical isolation: 

yes, galvanically

Temperature:

Range: 0 to 100°C
Resolution: 0.1°C
Accuracy: ±0.5°C
Sensor: 100 

or 1000 

Platinum RTD, 

terminal strip

Temperature compensation: 

Automatic/manual from –9.9 to 125°C

Temperature coefficient:

Conductivity: linear 0.00 to 10.00% per °C
Resistivity: ultra-pure/linear 0.00 to 

10.00% per °C

Calibration: 

Conductivity or resistivity: one point
Temperature: offset up to ±5°C

Password protection:

two level protection

with four-digit password lockout; 
first level gives calibration access, second
level allows set-up parameter changes

Display:

Dual-line LCD; 

4-digit upper and 3

1

2

-digit lower

Operating temperature: 

14 to 140°F 

(–10 to 60°C)

Housing:

IP54 front; 

1

4

DIN size 

Dimensions:  

Unit only: 3

13

16

"W x 3

13

16

"H x 6

15

16

"D 

(9.6 x 9.6 x 17.5 cm)

Boxed: 6

1

2

"W x 6"H x 9

1

4

"D 

(16.5 x 15.2 x 23.5 cm) 

Panel cut-out: 

3

3

8

"W x 3

3

8

"H (9.2 x 9.2 cm)

Weight: 

Unit only: 1.5 lbs (0.7 kg)
Boxed: 2.5 lbs (1.2 kg)

Summary of Contents for 1000 series

Page 1: ...ine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 17 Return of items Authorization must be obtained from our Customer Servi...

Page 2: ...on 33 6 3 4 Setting process temperature 33 6 3 5 Setting calibration temperature 34 6 4 SP1 SP2 Set up for Relay 1 and Relay 2 subgroup 35 40 6 4 1 Selecting relay set point values 36 6 4 2 Selecting...

Page 3: ...your solution s true temperature correction for higher accuracy Choice of 0 01 0 1 1 0 or 10 0 cell constants for accurate readings in any solution 0 20 4 20 mA transmitter recorder output for remote...

Page 4: ...l connector 2 2 2 3 Pt100 RELAY1 7 6 L N 5 4 PE 100mA F FUSE250VAC 11 HOLD ALARM RELAY2 1 2 3 17 17 16 15 14 12 13 Pt1000 18 NC 10 9 8 19 20 21 22 cell Flat Gasket 1 mm thick Installed by customer tap...

Page 5: ...lay common 9 High set relay working position 10 Alarm relay resting position IMPORTANT The alarm relay functions as an Active Low device e g it switches OFF under alarm condition Therefore the alarm d...

Page 6: ...and keys Let you 1 Select various parameter groups 2 Set parameters and numerical values in the Set up mode If you continuously hold the button the setting speed increases 3 Control the relays in the...

Page 7: ...e number see page 45 for more information C mS ATC READY MEAS CAL SETUP HOLD ERR m LED Indicators The LED indicators provide a quick way to check controller status They show whether a parameter within...

Page 8: ...cally reverts to Measurement mode if you do not press a button for 15 minutes Viewing operating parameters You can view operating parameters without entering the security code However you must enter t...

Page 9: ...n instructions A B A B Entering Advanced Set up mode 4 4 1 Press the ENTER key The upper display will show 000 and the lower display will show S Cd See figure 2 Press the key to scroll the upper displ...

Page 10: ...the value of your cell constant by using a standard conductivity calibration solution or a precision resistor for resistivity controllers You only need to know one of these values to calibrate your co...

Page 11: ...ibration solution so you will know the the solution s exact value 5 Press the and keys to match the display reading to the exact value of your conductivity solution See figure 6 Press the ENTER key to...

Page 12: ...23 6 Advanced set up mode The OAKTON 1000 series conductivity or resistivity controller features eight sub groups that organize all set up parameters These parameters let you customize your controlle...

Page 13: ...output Select value equivalent to 0 4 mA Select value equivalent to 20 mA CAL Con Res Calibration Calibrate by cell constant or by calibration standard See pages 26 29 for complete temperature compens...

Page 14: ...values use a linear temperature compensation curve for all other solutions Default is linear temperature compensation curve From measurement mode 1 Enter Advanced set up mode Push the ENTER key Press...

Page 15: ...see Appendix 2 on page 64 From measurement mode 1 Enter Advanced set up mode Push the ENTER key Press the key to scroll to security code 22 Push the ENTER key again 2 Press the and keys to scroll unti...

Page 16: ...ETUP HOLD SETUP HOLD 6 3 1 Selecting automatic or manual temperature compensation From measurement mode 1 Enter Advanced set up mode Push the ENTER key Press the key to scroll to security code 22 Push...

Page 17: ...r use a probe without a built in temperature sensor The controller will compensate for temperature at the values you enter For manual temperature compensation you can set two different temperatures pr...

Page 18: ...rn to Measurement mode by pressing the and keys escape simultaneously SP1 SP2 Set up for Relay 1 2 sub group SP1 SP2 Set up for relay SP1 SP2 overview SETUP HOLD mS SETUP HOLD SETUP HOLD mS SETUP HOLD...

Page 19: ...value for Set Point 1 Set Point 2 Your controller will activate at the value you select 5 Press the ENTER key to confirm your selection 6 Proceed to step 3 of section 6 4 2 or press the and keys toge...

Page 20: ...d to step 3 of section 6 4 4 or press the and keys together to return to measurement mode 6 4 4 Setting an on delay time lag NOTE this parameter appears only if on off control type is selected SETUP H...

Page 21: ...ENTER key again 2 Press the and keys to scroll until the upper display shows SP1 SP2 3 Press the ENTER key until the upper display shows off delay time and the lower display shows OF d See bold figur...

Page 22: ...ting break make contact relay type NOTE If the controller type LCt oFF is set this parameter is blanked out This mode lets you determine which relay state will trigger the alarm state de energized bre...

Page 23: ...ants K in ten possible combinations 1 0 000 to 2 000 S K 0 01 2 0 00 to 20 00 S K 0 01 3 0 00 to 20 00 S K 0 10 4 0 0 to 200 0 S K 0 10 5 0 0 to 200 0 S K 1 00 6 0 to 2000 S K 1 00 7 0 to 5000 S K 1 0...

Page 24: ...together to return to measurement mode SETUP HOLD mS SETUP HOLD SETUP HOLD mS SETUP HOLD mS SETUP HOLD SETUP HOLD mS SETUP HOLD SETUP HOLD mS SETUP HOLD mS SETUP HOLD 6 6 2 Choosing output range 4 20...

Page 25: ...ctivity resistivity value and the lower display shows r 20 See bold figures at right 4 Press the and keys to select the conductivity resistivity value equivalent to 20 mA 5 Press the ENTER key to conf...

Page 26: ...to measurement mode SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD 6 7 2 Selecting steady or pulse contact for the alarm relay This param...

Page 27: ...surement mode SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD 6 7 4 Reverting to factory default settings Use this parameter to reset all settings to factory default Changing from no to YES and...

Page 28: ...the controller will automatically return to the Measurement mode If you enter the Calibration mode from the Advanced Set up mode the controller will return to the sub group menu 6 8 For conductivity...

Page 29: ...TUP HOLD CAL HOLD B C S ATC CAL HOLD C C S ATC CAL HOLD D CAL HOLD E A A B C D E 1 Disconnect controller from power 2 Connect the resistor to the terminals used for the resistivity cell pins 21 and 22...

Page 30: ...lay mA 1 2 3 AC L N PE 4 5 6 7 8 9 10 11 12 13 14 15 16 17 Signal Output Hold Input Alarm Relay 2 Relay 1 Power Mains 10 Alarm relay resting position 11 Alarm relay common 12 Alarm relay working posit...

Page 31: ...y to return to Measurement mode The relays are now back under automatic control 10 Technical Specifications Range no Range Resolution Cell Constant K Conductivity controllers 1 0 000 to 2 000 S 0 001...

Page 32: ...xtra Controllers WD 35200 20 Conductivity controller 120 VAC 48 to 62 Hz WD 35200 25 Conductivity controller 220 VAC 48 to 62 Hz WD 35200 40 Resistivity controller 120 VAC 48 to 62 Hz WD 35200 45 Resi...

Page 33: ...mperature coefficient of your solution according to the formula shown above 5 Enter the temperature coefficient you calculated into the meter A From measurement mode Enter Advanced set up mode Push th...

Page 34: ...e lag 50 Setting steady or pulse contact 51 Assembly controller 5 10 ATC on or off selection 31 B Back panel 7 Break make contact relay type selecting 43 C Calibration conductivity Entering calibratio...

Reviews: