![Oakton 1000 series Operating Instructions Manual Download Page 24](http://html2.mh-extra.com/html/oakton/1000-series/1000-series_operating-instructions-manual_3107195024.webp)
46
47
6.6.3 Selecting conductivity (resistivity) value at 0 (4) mA
This parameter lets you choose a
conductivity (resistivity) value to be
equivalent to 0 (4) mA. Select the
lowest value of the conductivity
(resistivity) range that you want a
recorder or transmitter to detect.
From measurement mode:
1.
Enter Advanced set-up mode.
Push the ENTER key. Press the
▲
key to scroll to security code “22”.
Push the ENTER key again.
2.
Press the
▲
and
▼
keys
to scroll
until the upper display shows
“rng”.
3.
Press the ENTER key
until the
upper display shows a conductivity
(resistivity) value and the lower
display shows “r.0” or “r.4”
See bold figure at right.
4.
Press the
▲
and
▼
keys
to select
the conductivity (resistivity) value
equivalent to 0 (4) mA.
5.
Press the ENTER key
to confirm
your selection.
6.
Proceed to step 3 of section 6.6.4,
or press the
▲
and
▼
keys together
to return to measurement mode.
SETUP
HOLD
mS
SETUP
HOLD
SETUP
HOLD
mS
SETUP
HOLD
mS
SETUP
HOLD
SETUP
HOLD
mS
SETUP
HOLD
SETUP
HOLD
mS
SETUP
HOLD
mS
SETUP
HOLD
6.6.2 Choosing output range (4-20 or 0-20 mA)
This parameter lets you select an
output range of 4-20 or 0-20 mA. If you
select 4-20 mA, the minimum current
supplied by terminal 17 and 16 of the
17-way terminal block corresponds to
4 mA. If you select 0-20 mA, the
minimum current corresponds to 0.
From measurement mode:
1.
Enter Advanced set-up mode.
Push the ENTER key. Press the
▲
key to scroll to security code “22”.
Push the ENTER key again.
2.
Press the
▲
and
▼
keys
to scroll
until the upper display shows
“rng”.
3.
Press the ENTER key
until the
upper display shows the output type
(0-20 or 4-20) and the lower display
shows “out”.
See bold figure at right.
4.
Press the
▲
and
▼
keys
to select
your output range: 0-20 or 4-20 mA.
5.
Press the ENTER key
to confirm
your selection.
6.
Proceed to step 3 of section 6.6.3,
or press the
▲
and
▼
keys together
to return to measurement mode.