indicator enters setting / calibrating state. 

b. Need to open the indicator front-cover. 

      For entering setup/calibration, open its front cover, find out the short circuiter W1 and join 2 and 3 pin 

short. At this time the indicator automatically enter setup/calibration mode and displays  “CAL    0” 

after  “SET    UP”  is  displayed  for  about  2  seconds.  Now  the  indicator  enters  setting  /  calibrating 

state. 

Note:

 Under setup/calibration mode, you can set the data by using the following keys. 

[↑]:          used for increasing the value of the selected place. 

[ENTER]:used for confirming the selected value and entering the next-grade data setup.     

[→]:        used for selecting the setup mode or selecting the given data-table or shifting place while 

inputting the weight value on arbitrary point. 

8.2 Entering calibration or ambient data group 

“CAL    X”    refers to fixed point calibration / arbitrary point calibration /no calibration, you can use 

[→] to select the calibration mode, the method are as follows:     

“X=0” refers to no calibration,press [ENTER] to enter the ambient data group 

“X=1” refers to fixed point calibration,press [ENTER] to enter the indicator calibration (range fixing). 

“X=2” refers to arbitrary point calibration,press [ENTER] to enter the indicator calibration (range 

arbitrary point). 

Note:

 the weight value of the calibration weights is 5%Wmax<Wadd<Wmax,Wmax refers to full 

range and Wadd refers to the loading-weights value.     

8.

3 Calibration process 

Step 1: full range and division selecting 

       

Use 

[

to select the full range value to be calibrated, then press [ENTER] to confirm. At this time 

the indicator enter division selecting, also press [ENTER] to confirm after selecting division. 

Summary of Contents for HL-2065

Page 1: ...XK3101 HL 2065 Indicator User s Manual...

Page 2: ...uction 2 4 Installation Connection 4 5 Operation Keys Instruction 4 6 Use Instruction 5 6 1 Turn on Operation 5 6 2 Turn off Operation 5 6 3 Zeroing 5 6 4 Deducting Tare 5 6 5 Conversion between kg an...

Page 3: ...drift 12 ppm 1 4 A D conversion resolution ration 300000 1 5 Display resolution ration 30000 1 6 A D conversion rate 100T S 1 7 Display shift rate 5T S 1 8 Non linearity 0 01 F S 1 9 Load cell excitin...

Page 4: ...101 HL 2065C 1 5kg 2 Main Functions 2 1 Full keyboard calibrating 2 2 Zero position auto tracking 2 3 Full range deducting tare 2 4 Over load indicating 2 5 Simple counting 2 6 Two kinds of calibratio...

Page 5: ...4 Figure 1 XK3101 HL 2065B Figure 2 XK3101 HL 2065C 3 2 Panel...

Page 6: ...icating light MODE key TARE ENTER key ZERO key PRINT ON OFF key 3 3 Cursors and their meanings There are 6 LED indicating cursors on this indicator for the names and meanings see the following table C...

Page 7: ...ween load cell and 5 core aerial pin outlet 4 2 2 If the 9 core D shape plug is used for the load cell connecting see Figure 4c Figure 4b Wire connecting between load cell and 9 core D shape plug Note...

Page 8: ...print signal MODE key a Used for converting the mode between weighing and counting b Under calibrating state you can use this key to choose the setup method or the given parameter table also use it to...

Page 9: ...goods weight is 0 5kg can be regarded as tare Press TARE 0 00 The tare has been deducted and at this time the NET indicating cursor appears Remove the goods 0 50 The indicator displays the tare negat...

Page 10: ...t the same time 6 7 RS 232 Output 6 7 1 The 3 core aerial outlet is used for RS 232 hardware interface and the outlet is mounted on the back housing of the indicator see the following drawing Figure 5...

Page 11: ...s weight D0 1 net weight State C output empty case character 6 7 3 Every frame consists of 10 digits The first is start bit 0 The ninth is stop bit 1 One of the other eight is correction bit and the o...

Page 12: ...s used for selecting digit place and the digit of the selected place will flash at this time if you press the flashing digit of the selected place will be changed in 1 2 3 9 0 1 sequence For instance...

Page 13: ...ct the setup value of the products leaving the company 7 6 Turn off its power at once if the indicator is out of order in use Non specialized manufacturer or non professional staff can t fix it withou...

Page 14: ...he weight value on arbitrary point 8 2 Entering calibration or ambient data group CAL X refers to fixed point calibration arbitrary point calibration no calibration you can use to select the calibrati...

Page 15: ...IV data for about 2 seconds At this time you can press to select the division and then press ENTER to confirm After the operation is finished the indicator enters zero load adjustment state For the ra...

Page 16: ...Ad LD Load the weights on the scale platform after the indicator display Ad LD for 2 seconds Await 10 seconds make sure that the indicator is fully stable then press ENTER to confirm at this time the...

Page 17: ...etup F2 X refers to auto zero tracking range X 0 5 1 0 2 0 3 0 d The auto zero tracking function can auto compensate the zero point change caused by the factor that a little material to be weighed dro...

Page 18: ...e wave filtering strength the quicker the indicator response speed and the less stably the indicator works On the other hand the higher the wave filtering strength the slower the indicator response sp...

Page 19: ...he indicator enters F9 setup F9 X refers to user s setup confirming or return to factory setup data X 0 confirm the user s setup data X 1 return to the factory setup data See the following table for t...

Page 20: ...icator displays 0 000 00 the flashing 0 at the first digit on the last left is the selected digit Load 20kg rectification weights on the scale platform press twice to select the digit input place the...

Reviews: