13
indicator enters setting / calibrating state.
b. Need to open the indicator front-cover.
For entering setup/calibration, open its front cover, find out the short circuiter W1 and join 2 and 3 pin
short. At this time the indicator automatically enter setup/calibration mode and displays “CAL 0”
after “SET UP” is displayed for about 2 seconds. Now the indicator enters setting / calibrating
state.
Note:
Under setup/calibration mode, you can set the data by using the following keys.
[↑]: used for increasing the value of the selected place.
[ENTER]:used for confirming the selected value and entering the next-grade data setup.
[→]: used for selecting the setup mode or selecting the given data-table or shifting place while
inputting the weight value on arbitrary point.
8.2 Entering calibration or ambient data group
“CAL X” refers to fixed point calibration / arbitrary point calibration /no calibration, you can use
[→] to select the calibration mode, the method are as follows:
“X=0” refers to no calibration,press [ENTER] to enter the ambient data group
“X=1” refers to fixed point calibration,press [ENTER] to enter the indicator calibration (range fixing).
“X=2” refers to arbitrary point calibration,press [ENTER] to enter the indicator calibration (range
arbitrary point).
Note:
the weight value of the calibration weights is 5%Wmax<Wadd<Wmax,Wmax refers to full
range and Wadd refers to the loading-weights value.
8.
3 Calibration process
Step 1: full range and division selecting
Use
[
→
]
to select the full range value to be calibrated, then press [ENTER] to confirm. At this time
the indicator enter division selecting, also press [ENTER] to confirm after selecting division.