12
Step 7:
Press [TARE] to confirm. Thus the input operation for a group of digit is completed.
7.
Use Guide
7.1
Be sure that the ground terminal of AC power is properly grounded.
7.2
Weighing (tare included) can’t exceed Max. Capacity plus 9e (“e” refers to graduation). If exceed,
the indicator displays “OVER”.
7.3
Don’t disassemble it without authorization.
7.4
After turn-on, if the indicator doesn’t display or its display don’t change, at this time you can try to
turn off its power and turn on its power again.
7.5
When you set “F1” data, you may have not enough patients or make a mistake. If you don’t intend
to be a more professional personnel, you only need to make
F7= 1,
thus
you must not consider
selecting the other data of F1~F6 because at this time the indicator (scale) can automatically
select the setup value of the products leaving the company.
7.6
Turn off its power at once if the indicator is out of order in use. Non-specialized manufacturer or
non-professional staff can’t fix it without authorization. Users can contact our selling agent in your
area.
Caution:
Never change its data presumptuously once when the indicator can normally be
operated, unless the act is permitted by the laws.
7.7
If you need to re-calibrate the indicator, do it after the indicator is turned on for 30 minutes in order
to ensure its circuit in stable working state.
8. Calibration
8.1 Calibrating and data setting
There are 2 methods for you to enter setting / calibrating mode.
a. Don’t need to open the indicator front-cover.
After power-on, keep pressing [MODE] and [ZERO] while the indicator is self-checking “0 ~ 9”,
loose the 2 keys at the same time if the indicator displays “Und0”. At this time the indicator displays
“CAL X” (“X” refers to 0 or 1 or 2) after “SET UP” is displayed for about 2 seconds. Now the