4.2
ADJUSTING THE LAMP POSITION
Be sure that the large upper door of the lamp housing is closed before you adjust
the lamp position. The lamp positioning should be adjusted to achieve maximum
output. With a lamp warmed-up to normal operating temperature, removed the
small door of the lamp main housing. Open the shutter, place a reference cell
measurement probe at the center of the exposure area, and vary the adjustments
for maximum intensity. Due to subtle differences between the lamps, the X, Y,
and Z adjustment should be a readjusted every time a lamp is replaced. This is
referred to as "peaking" a lamp.
XYZ lamp position adjustment knobs shown with lower panel removed.
4.3 CLEANING OPTICAL COMPONENTS
Periodic cleaning of optics is recommended to maintain light intensity and
uniformity. Items to be cleaned include the mirrors and lens. Do not attempt to
clean the condensing lens array (integrator), elliptical reflector or feedback
probes. These should only be disassembled and cleaned at OAI.
With the light source fully cooled, gently clean the optics with a soft, lint-free,
lens tissue and a small amount of isopropyl alcohol.
NOTE
Dispose of the alcohol soaked lens tissue as described in Section 7
HAZARDOUS MATERIALS USED WITH THE SYSTEM