D 8.4.1.20 O
2
matic User Manual - IFU
70
Table A5
–
Factory set profiles
Name
Min
SpO2
Max
SpO2
Min
O2
Flow
Max
O2
Flow
Initial
O2
Flow
Max
Pulse
rate
Flow
response
COVID_LFLOW
92
96
0
8
2
130
SLOW
COVID_HFLOW
92
96
0
15
5
130
NORMAL
COPD_NORM
88
92
0
6
2
130
SLOW
ASTHM_PNEUM
94
98
0
10
3
130
SLOW
WALKING
90
94
0
15
2
150
FAST
Flow response
Rate
Slow
25 %
Normal
100 %
Fast
150 %