D 8.4.1.20 O
2
matic User Manual - IFU
38
Figure 14: O
2
matic PRO 100 trend screen
–
last 8 hours
In Figure 15, an extract of the trend screen is shown.
Legends can comprise of different compositions.
Either it can be a solid vertical line, a dotted vertical line, a
downward facing triangle or a combination of the lines and triangle.
The solid vertical line indicates that the device has been rebooted or
turned off and on.
The dotted vertical line indicates that the patient has been altered or
changed.
The downward facing triangle indicates that the profile has been
altered or changed to another profile.
Figure 15: Extract from figure 14 focusing on the legends
Info screen
The Info screen shows average SpO
2
, O
2
and pulse rate numbers for
the last 24 hours, as well as minimum (MIN) and maximum (MAX)
for each treatment parameter. The T-HH:MM at MIN and MAX show
how long ago the MIN or MAX occurred. For example, in below info
Summary of Contents for OMC PC-100
Page 1: ...2022 6 09 ...
Page 16: ...D 8 4 1 20 O2matic User Manual IFU 16 ...
Page 18: ...D 8 4 1 20 O2matic User Manual IFU 18 ...
Page 24: ...D 8 4 1 20 O2matic User Manual IFU 24 ...
Page 26: ...D 8 4 1 20 O2matic User Manual IFU 26 ...
Page 27: ...D 8 4 1 20 O2matic User Manual IFU 27 ...
Page 28: ...D 8 4 1 20 O2matic User Manual IFU 28 ...
Page 29: ...D 8 4 1 20 O2matic User Manual IFU 29 ...
Page 36: ...D 8 4 1 20 O2matic User Manual IFU 36 ...
Page 41: ...D 8 4 1 20 O2matic User Manual IFU 41 Figure 17 SpO2 outside safe range prompt ...
Page 74: ...D 8 4 1 20 O2matic User Manual IFU 74 Notes ...
Page 75: ...D 8 4 1 20 O2matic User Manual IFU 75 Notes ...