D 8.4.1.20 O
2
matic User Manual - IFU
7
“ON”
“OFF”
Class II
equipment
IPX0
Ingress
protection
from dust
and water
Equipment shall
only be used
with
rechargeable
battery.
WARNING
1.6
Essential performance
The essential performance of the
O
2
matic PRO 100
device is to measure
pulse and SpO
2
, to maintain or adjust oxygen flow to the patient in response
to changes to measured values according to the user defined target values
and range thresholds. If the measured values are outside the user specified
limits, the O
2
matic PRO 100 device shall trigger the relevant alarms.
1.7
Operators and safety
O
2
matic PRO 100 should only be used by medical staff, with basic
experience in the use of pulse oximeters and treatment of patients in need
of oxygen therapy. It is recommended that each hospital appoints
administrators responsible for setting up the device metadata, i.e. standard
profiles and alarm delays. These should normally be physicians.
The O
2
matic PRO 100 has a failsafe rechargeable battery that keeps the
device running upon short power breaks. The device has two power related
buttons,
Main power
and
Standby,
for the purpose of distinguishing
between a power break and a proper shutdown. Once the device has been
turned on, a proper shut down is needed for the device not to trigger any
alarms. To shut down O
2
matic PRO 100 press and hold
(A)
for 3
seconds.
Summary of Contents for OMC PC-100
Page 1: ...2022 6 09 ...
Page 16: ...D 8 4 1 20 O2matic User Manual IFU 16 ...
Page 18: ...D 8 4 1 20 O2matic User Manual IFU 18 ...
Page 24: ...D 8 4 1 20 O2matic User Manual IFU 24 ...
Page 26: ...D 8 4 1 20 O2matic User Manual IFU 26 ...
Page 27: ...D 8 4 1 20 O2matic User Manual IFU 27 ...
Page 28: ...D 8 4 1 20 O2matic User Manual IFU 28 ...
Page 29: ...D 8 4 1 20 O2matic User Manual IFU 29 ...
Page 36: ...D 8 4 1 20 O2matic User Manual IFU 36 ...
Page 41: ...D 8 4 1 20 O2matic User Manual IFU 41 Figure 17 SpO2 outside safe range prompt ...
Page 74: ...D 8 4 1 20 O2matic User Manual IFU 74 Notes ...
Page 75: ...D 8 4 1 20 O2matic User Manual IFU 75 Notes ...