37
Maintenance & Service
The AUTOCLEAVER
TM
is designed for a production environment and to give trouble-free operation provided normal
maintenance is performed. Improper service and/or repair could result in the safety features being disabled and can also
lead to damage that will not be covered under warranty. Therefore NYFORS recommends that maintenance and service
procedures should be performed only by trained personnel.
Maintenance Schedule
As needed
Daily
Monthly
12 months
Inspect/Clean v-groove and fixed clamp block
X
X
Inspect/Clean v-groove slide ramp
X
X
Inspect/Clean height adjustor slide ramp
X
Empty fiber waste container
X
X
Clean cleaver blade
X
X
Repositioning cleaver blade
X
Cleaver blade replacement
X
Verify/Recalibrate load cell
X
Inspection & Cleaning
General cleaning
•
Isopropyl alcohol (IPA) is the recommended cleaning solvent.
•
Never use spray cleaner or compressed gas for cleaning the AUTOCLEAVER
TM
.
•
Never pour liquid cleaner on the instrument. Always use a clean applicator to apply the solvent and clean the
exposed parts.
•
Never use hard objects, acetone, thinner or benzol to clean the AUTOCLEAVER
TM
.