18 Inspection plan
System Data
Supplier:
Date of installation:
Nuvo serial no.:
Operator's serial no.:
Device location:
Important information
• The inspection intervals must be observed.
• This inspection plan is only valid when combined with the Nuvo Installation and Operating Instructions which must be consulted as
complementary reference documents during inspections.
The pendant system must be checked by Nuvo or a company authorised by Nuvo for the following points at the intervals specified
below:
Visual inspection (annually)
NOK
OK
• The parts of the pendant system are not deformed and are free from damage (scratches, cracks, etc.)*
• The system is free from defects in paintwork*
• The plastic covers are available and correctly positioned*
• All ratings plates and labels are available and readable
• All the plastic screws of the canopy are available and fixedly attached
• All cables and gas hoses must be inspected for damage and replaced if required**.
Nuvo recommends that you replace cables and gas hoses after 10 years.
Functional check (annually)
NOK
OK
• The system is safely positioned; adjust the limit stops if required
• At least 1 limit stop is set at each centre of rotation, properly attached and effective**
• After releasing the pneumatic brake or electromagnetic brake, the extension arms can be conveniently positioned
• The brake valves have been vented & close properly; the brake buttons operate smoothly***
• The mechanical brakes (friction brakes) have been adjusted properly; re-adjust them if required
• The system remains in any selected position**
• All sockets (if existing) have been fixedly mounted and function properly**
• Protective conductor resistance checked** (DIN EN 62353)
• The screw type connections, with the torques specified in the Installation Instructions, are securely in place
• The gas sockets (if existing) have been fixedly mounted and function properly**
• Inspection of the gas hoses/sockets (if exist.) by an authorised body with regard to the gas type/gas density**
• Load compensation/spring tension correct, re-adjust it if req.**
• The BrakeGuides (optional) light up when pressing the corresponding brake button*
• The extension arm lighting (optional) lights up when actuating the switch (on the Navigator M6 / CEMOR)**
Remarks
Confirmation of the inspection
The work mentioned above has been executed, including the adjustments required and the safety test:
Date
Name (in block letters)
Signature/Stamp
* Damaged, deformed or missing components must be replaced as a precaution. In this case, contact the supplier of the pendant system.
** If one of the points mentioned above is found to be incompliant during the inspection, the pendant system must be placed out of operation immediately as a precaution in
order to prevent further damage to persons and equipment. Notify the system supplier immediately.
*** Only applies to Navigator Lift
™
180 Air
Plus
The Medical Product Book which belongs to the medical product in accordance with Medical Devices Operators Ordinance (MPBetreibV) must be available on site. Service
and maintenance work and security checks must be documented in this Medical Product Book. Test reports, including this report, must be filed out in the corresponding
Medical Product Book.
Summary of Contents for Navigator Lift 180 Air Plus-Single