14 Electromagnetic Compatibility (EMC) Information
en
Cont.
The Navigator Lift
™
180 is intended for use in the ELECTROMAGNETIC environment
specified below. The customer or the user of the Navigator Lift
™
180 should ensure that it
is used in such an environment.
Interference immunity test
Test level in accordance with IEC 60601
Compliance level
Immunity to conducted disturbances,
induced by radiofrequency fields
IEC 61000-4-6
3V
0.15MHz to 80MHz
6V in ISM frequency bands from 0.15MHz
to 80MHz
80% AM at 1kHz
passed
High-frequency electromagnetic fields in ac
-
cordance with
IEC 61000-4-3
3V/m
80MHz to 2.7GHz
80% AM at 1kHz
passed
NOTE 1
At 80 MHz and 800 MHz, the higher value applies.
NOTE 2
These guidelines may not apply in all cases. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
a
The field strength of stationary transmitters, including the base stations of mobile phones and mobile land mobile radios, amateur radio
stations, AM and FM radio and TV broadcasting transmitters, cannot be precisely predetermined theoretically. To assess the electromag
-
netic environment due to stationary RF transmitters, an electromagnetic site survey should be considered. If the field strength measured
in the location in which the Navigator Lift
™
180 is used exceeds the applicable RF compliance level above, the Navigator Lift
™
180 should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the Navigator Lift
™
180.
b
Field strengths over the 150 kHz to 80 MHz frequency range should be less than 3 V/m.