Never use this device when ambient temperature is over 40°C. For more operating 

environment requirements, please refer to the section - Specifications.
Do not shake the nebulizer during operation.
Avoiding any strong shock on the main unit and its components, e.g., falling on the 

ground.
Never dismantle the device or attempt to repair any part or component.
This device is for human use only.
Always use optional accessories as stated in this Instruction Manual only.
The device and its parts and components and optional accessories shall only be 

discarded in accordance with the relevant local regulations. Inappropriate disposal 

may cause environmental pollution.
Prior to use, make sure the device and its parts and components (e.g. mouthpiece 

and mask) are correctly assembled as per this Instruction Manual.

Products with Mesh technology, are not able to nebulise drugs with a high molecular 

density such as drugs based on hyaluronic acid.  To preserve the pores of the nebulizing 

plate and to avoid clogging it is advisable to avoid the use of physiological hypertonic 

solution due to the high salt concentration. For this reason it is better to dilute the 

drugs ONLY with isotonic solution.

Never use the device in case the power cord is wet.
Never plug in or unplug from the power socket with wet hands.
Do not immerse the main unit into water or other liquids.
Protect it against water or other liquid splashes. In case it contacts with any splash, 

disconnect the power supply and mop up with gauze or other soft adsorbing 

materials.
Never use the device in humid environment such as bathroom.
When connected to power supply, check the supply voltage at power socket is 

correct and make sure your plugging in will not cause overload.
Disconnect it from power supply after the use of this device and never leave an 

energized device aside.
Clean the device only after disconnecting it from power supply.

• 


• 




• 

 

• 


• 

 


• 

• 

Caution:

Risk of electric shock:

SAFETY INFORMATION

WARNINGS

Summary of Contents for 5076

Page 1: ...EN Portable nebuliser with MESH technology 5076...

Page 2: ...Manufacturer Low battery indicate Power indicate light Keep away from sunlight Type BF Applied Part Read instructions carefully Keep in a cool dry place Operating instructions Symbol for the marking o...

Page 3: ...molecular density such as drugs based on hyaluronic acid To preserve the pores of the nebulizing plate and to avoid clogging it is advisable to avoid the use of physiological hypertonic solution due t...

Page 4: ...9 10 11 12 13 14 a b 15 Adult Mask Child Mask Adapter Optional Power Plug Adapter Connector Storage bag PREPARATION BEFORE USE Use the Battery Open the battery cover in the bottom of the main unit al...

Page 5: ...ning and disinfecting To avoid cross infection all users of a same device should each be provided with a set of nebulizer and accessories The nebulizers and accessories for different users should be s...

Page 6: ...left hand and close the nebuliser cover with your right hand press the lock downwards with the thumb to the point when you hear a clicking sound that indicates a good closing state 5 Reinstall the Ou...

Page 7: ...sal cavity it is recommended to take a long breath through nasal cavity and then exhale through mouth In case of medication through throat it is recommended to take breath through mouth and then exhal...

Page 8: ...nds depending on the liquid type When there is not much liquid left you may tilt the device towards yourself to allow liquid gathered at the nebulising head side in order to make full use of the resid...

Page 9: ...ting the nebulising head gently clean the metal mesh and electrode in the nebulising head with a medical swab dipped with medical alcohol rinse the nebulising head with water dry the surface of nebuli...

Page 10: ...our doctor No tap water or boiled water can be used The container for preparation of liquid must be boiling sterilised The duration limit for each use of the device should be no more than 15 minutes o...

Page 11: ...vice does not work Incorrect installation of nebuliser Reinstall the nebuliser correctly Dirt on and in the nebuliser is serious Clean the nebuliser replace nebulising head if necessary The liquid can...

Page 12: ...y Life 90 minutes supplied by two AA Alkaline Battery Automatic Shutdown Automatic shutdown without liquid Operating Temperature and humidity Air pressure 10 C 40 C 85 R H 700hPa to 1060hPa Storage Te...

Page 13: ...d as consumable e g batteries brush heads or parts subject to usury The legal guarantee of 24 months is void if 1 2 3 The product has undergone aesthetic damage due to improper use not in accordance w...