NUVE NF 1200 User Manual Download Page 6

 

SECTION 2 

TECHNICAL SPECIFICATIONS 

2.1  TECHNICAL SPECIFICATIONS TABLE 

 

 

NF 1200 

NF 1200 R 

Max Speed 

RA 280 rotor: 4100 rpm, MP 200 rotor: 4100 rpm,  

RS 600 rotor: 9000 rpm, RS 100 rotor: 12000 rpm, RS24 rotor: 14000 rpm,  

RS 30 rotor: 13500 rpm 

Max. RCF 

RA 280 rotor: 3082 x g, MP 200 rotor: 2349 x g, RS 600 rotor: 8965 x g, RS 

100 rotor: 14167, RS24 rotor : 18405 x g, RS 30 rotor: 20170 x g 

Max. Capacity 

RA 280 rotor: 4x280 ml, MP 200 rotor: 2x3 microtitre plates,  

RS 600 rotor: 6x100 ml, RS 100 rotor: 10x10 ml, RS 24 rotor: 24x1.5/2 ml, 

RS 30 rotor: 30x2 ml 

Control System 

Microprocessor controlled 

Display 

High Visibility Digital 

No. of Memories 

10 

Speed Range 

500-14000 rpm 

Speed Step 

10 rpm 

Speed Accuracy 

± 20 rpm 

Timer Range 

01-99 min + Hold Position 

Timer Step 

1 min 

Acceleration Rate 

0: Slowest   9: Fastest 

Braking Rate 

0: Slowest   9: Fastest 

Power supply 

230 V, 50 Hz 

Temp. range 

 

-

9° C / 40 ° C 

Refrigerant Liquid 

 

R134a 

Temp. Accuracy 

 

1°C 

Temp. Performance 

(at maximum speed) ambiant temp+12°C 

+4°C at max. speed 

Max. Power 

500 W 

800 W 

Ext. Dimension 

455 x 545 x 400 

770 x 540 x 400 

Packing Dim. 

600 x 530 x 470 

870 x 650 x 550 

Net / Gross Weight 

42 / 47 kg 

75 / 80 kg 

 

2.2  ACCESSORIES FOR NF 1200/1200R

 

 

A 08 062 

Puller for rotor removal 

 

B 50 016 

RA 280 swing- out rotor 4*280 ml , 4.100rpm , 3.082 x g 

 

G 51 007 

Set of 4 buckes for RA 280 

 

G 51 036 

Set of 4 buckets with sealing cap 

 

G 51 008  

Set of 4 insert 1x280 ml 

 

G 51 009 

Set of 4 insert 1x250 ml flat  

 

G 51 010 

Set of 4 inserts 1x200 ml flat 

 

G 51 011 

Set of 4 inserts 1x100 ml 

 

G 51 012 

Set of 4 inserts 2x50 ml conical 

 

G 51 013 

Set of 4 inserts 4x25 ml 

 

G 51 014 

Set of 4 inserts 2x15 ml conical 

Summary of Contents for NF 1200

Page 1: ...0 N VE SANAY MALZEMELER MALAT VE T CARET A NF 1200 1200 R MULTI PURPOSE BENCH TOP CENTRIFUGES USER S MANUAL Z14 K25 249 Rev No 07 Rev Date 01 2016...

Page 2: ...lity in the case where the goods are not used in accordance with their proper intent 4 The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper hand...

Page 3: ...ove the warning labels do not operate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your N ve agent for...

Page 4: ...CTING THE MEMORIZED PROGRAM 13 4 5 SAFETY INTERLOCK SYSTEM 14 4 6 IMBALANCE DETECTION SYSTEM 14 4 7 MANUAL LID OPENING 14 SECTION 5 15 OPERATING PRINCIPLES 15 5 1 PREPARATION OF THE ROTOR TO RUN 15 5...

Page 5: ...h a cooling system which provides to operate between 9 C and 40 C It ensures safe processes by means of the locking system which does not allow the centrifuge to operate if the lid is open and which d...

Page 6: ...ge 01 99 min Hold Position Timer Step 1 min Acceleration Rate 0 Slowest 9 Fastest Braking Rate 0 Slowest 9 Fastest Power supply 230 V 50 Hz Temp range 9 C 40 C Refrigerant Liquid R134a Temp Accuracy 1...

Page 7: ...daptors 1x50 ml G 51 019 Set of 6 adaptors 1x50 ml cocinal G 51 020 Set of 6 adaptors 1x30 38 ml G 51 021 Set of 6 adaptors 1x15 ml cocinal B 50 022 RS 100 Angle rotor 10x10 ml 12 000 rpm 14 167xg G 5...

Page 8: ...mm Max Speed rpm Max RCF xg RA280 Swing out Rotor 4x280ml 164 4 100 3 082 Set of 4 buckets 380 ml nominal 76 Set of 4 inserts 1x280ml 65 Set of 4 inserts 1x250ml flat 62 Set of 4 inserts 1x200ml flat...

Page 9: ...ailure The working bench should be durable to the device weight and vibration isolated Ensure that the rotor is placed correctly prior to usage According to the standard IEC 61010 2 020 anyone and any...

Page 10: ...d motor shaft IF MENTIONED WARNINGS ARE NOT CONSIDERED N VE WILL NOT BE RESPONSIBLE FROM THEIR RESULTS SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol...

Page 11: ...y of the run time from 1 min to 99 min and hold position H 03 Acceleration Brake Temperature Display Display of the acceleration and brake rates from 0 to 10 for NF 1200 R it also displays the bowl te...

Page 12: ...Keys They increase or decrease values during programming stage and to select the program number during stand by position By pushing value increase key it is possible to see the set values during centr...

Page 13: ...on the time display 2 flashes Enter the run time by pushing the value increase and decrease keys 16 H means hold position Push the function key 16 and see that the acceleration led turns on and the v...

Page 14: ...tion led turns on The set speed or RCF value is displayed during the set time duration RPM or RCF screen will appear in every 3 seconds For the operations whose speeds are higher than 2 000 rpm the ro...

Page 15: ...electronically when an unacceptable imbalance occurs In this case the brake is applied immediately The Err 1 imbalance error is displayed and the imbalance indicator turns on The lid can not be opene...

Page 16: ...es should be at the same height to balance the load Dynamic Balance This balance states that diametrically oppositely replaced center of the gravity of tubes are symmetrical with respect to spinning a...

Page 17: ...t the rotor fits on the drive shaft See Figure 4 Screw the shaft nut with socket wrench to the clockwise direction Make sure that shaft nut is screwed tightly but do not expose over force to the socke...

Page 18: ...quently with the oil provided with the centrifuge Please remove the light oil from the pins and put a small amount of fresh oil every time you grease This will ensure free swinging of the buckets Most...

Page 19: ...e chemical the stress corrosion starts This type of corrosion is even more dangerous than the chemical corrosion as the effects of this corrosion are microscopic and very difficult to observe in the c...

Page 20: ...o the supply The electricity installation is not defective Power is supplied In case of below written failures related error codes are shown Err 2 Motor overheat Please switch off the centrifuge for a...

Page 21: ...20 SECTION 9 ELECTRICAL CIRCUIT DIAGRAMS 9 1 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200...

Page 22: ...21 9 2 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200R...

Page 23: ...22 SECTION 10 WARNING LABEL GROUNDED SOCKET NF 800R FUSES 2x10A NF 800 FUSES 2x4A...

Reviews: