NUVE NF 1200 User Manual Download Page 15

14 

 

NOTE  3 :

 

At  hold  position,  the  rotor  keeps  spinning  until  the  STOP  key  (7)  is  pushed. 
After  the  STOP  has  been  pushed,  the  rotor  brakes  according to  the  set  brake 
rate,  

NOTE  4 :

 

The  centrifuge  waits  at  the  stand-by  position  with  the  program  values  of  the 
last program. 

NOTE  5 :

 

Please  open  the  lid  after  every  operation,  otherwise  the  centrifuge  cannot  be 
run again. 

 

4.5.  SAFETY INTERLOCK SYSTEM 

 
The safety interlock system prevents opening of the lid when the rotor is spinning. 
The  centrifuge  does  not  operate  until  the  lid  is  closed  and  the  lid  remains  locked  until  the  rotor 
stops spinning. The 

“Lid Open” indicator turns on to warn the user if the lid is not closed properly.  

 

NOTE: 

If power failure occurs, access to the samples is possible by opening the lid with a special 
tool. Please see the manual lid opening section (3.6.) for further information. 

4.6.  IMBALANCE DETECTION SYSTEM 

  The  imbalance  detection  system  of  the  NF  1200  /  NF  1200  R  centrifuges  operates 

electronically  when  an  unacceptable  imbalance  occurs.  In  this  case  the  brake  is  applied 
immediately. 

 

The “Err 1” imbalance error is displayed and the imbalance indicator turns on. The lid can not 
be  opened  until  the  rotor  stops  spinning.  The  centrifuge  can  only  be  started  after  the  lid  has 
been opened and the rotor has been re-loaded correctly.  

  To avoid facing imbalance problem, please make sure to insert the tubes correctly. 

4.7.  MANUAL LID OPENING 

In  case  of  power  cut  or  any  breakdown,  the  centrifuge  can  be  opened  manually  to  access  the 
samples. 
 
To open the lid manually: 
 

  Switch off the instrument. 

  Insert  the  manual  lid  opening  tool  into  the  hole  on  the  left  side  of  the  instrument  (see  the 

general view drawing). 

  Push the tool keeping it horizontal until the lid opens. 

 

WARNING !!! 
 

Before opening the lid manually, make sure that the rotor already stopped spinning. Upon opening 
the  lid,  lift  it  by  hand  and  observe  the  rotor.  If  the  rotor  is  still  spinning,  close  the  lid  and  wait 
approximately  10  minutes  before  repeating  the  operation.  This  operation  must  be  carried  by 
someone who is informed of the danger and of the precautions which must be undertaken.

 

 

 

Summary of Contents for NF 1200

Page 1: ...0 N VE SANAY MALZEMELER MALAT VE T CARET A NF 1200 1200 R MULTI PURPOSE BENCH TOP CENTRIFUGES USER S MANUAL Z14 K25 249 Rev No 07 Rev Date 01 2016...

Page 2: ...lity in the case where the goods are not used in accordance with their proper intent 4 The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper hand...

Page 3: ...ove the warning labels do not operate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your N ve agent for...

Page 4: ...CTING THE MEMORIZED PROGRAM 13 4 5 SAFETY INTERLOCK SYSTEM 14 4 6 IMBALANCE DETECTION SYSTEM 14 4 7 MANUAL LID OPENING 14 SECTION 5 15 OPERATING PRINCIPLES 15 5 1 PREPARATION OF THE ROTOR TO RUN 15 5...

Page 5: ...h a cooling system which provides to operate between 9 C and 40 C It ensures safe processes by means of the locking system which does not allow the centrifuge to operate if the lid is open and which d...

Page 6: ...ge 01 99 min Hold Position Timer Step 1 min Acceleration Rate 0 Slowest 9 Fastest Braking Rate 0 Slowest 9 Fastest Power supply 230 V 50 Hz Temp range 9 C 40 C Refrigerant Liquid R134a Temp Accuracy 1...

Page 7: ...daptors 1x50 ml G 51 019 Set of 6 adaptors 1x50 ml cocinal G 51 020 Set of 6 adaptors 1x30 38 ml G 51 021 Set of 6 adaptors 1x15 ml cocinal B 50 022 RS 100 Angle rotor 10x10 ml 12 000 rpm 14 167xg G 5...

Page 8: ...mm Max Speed rpm Max RCF xg RA280 Swing out Rotor 4x280ml 164 4 100 3 082 Set of 4 buckets 380 ml nominal 76 Set of 4 inserts 1x280ml 65 Set of 4 inserts 1x250ml flat 62 Set of 4 inserts 1x200ml flat...

Page 9: ...ailure The working bench should be durable to the device weight and vibration isolated Ensure that the rotor is placed correctly prior to usage According to the standard IEC 61010 2 020 anyone and any...

Page 10: ...d motor shaft IF MENTIONED WARNINGS ARE NOT CONSIDERED N VE WILL NOT BE RESPONSIBLE FROM THEIR RESULTS SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol...

Page 11: ...y of the run time from 1 min to 99 min and hold position H 03 Acceleration Brake Temperature Display Display of the acceleration and brake rates from 0 to 10 for NF 1200 R it also displays the bowl te...

Page 12: ...Keys They increase or decrease values during programming stage and to select the program number during stand by position By pushing value increase key it is possible to see the set values during centr...

Page 13: ...on the time display 2 flashes Enter the run time by pushing the value increase and decrease keys 16 H means hold position Push the function key 16 and see that the acceleration led turns on and the v...

Page 14: ...tion led turns on The set speed or RCF value is displayed during the set time duration RPM or RCF screen will appear in every 3 seconds For the operations whose speeds are higher than 2 000 rpm the ro...

Page 15: ...electronically when an unacceptable imbalance occurs In this case the brake is applied immediately The Err 1 imbalance error is displayed and the imbalance indicator turns on The lid can not be opene...

Page 16: ...es should be at the same height to balance the load Dynamic Balance This balance states that diametrically oppositely replaced center of the gravity of tubes are symmetrical with respect to spinning a...

Page 17: ...t the rotor fits on the drive shaft See Figure 4 Screw the shaft nut with socket wrench to the clockwise direction Make sure that shaft nut is screwed tightly but do not expose over force to the socke...

Page 18: ...quently with the oil provided with the centrifuge Please remove the light oil from the pins and put a small amount of fresh oil every time you grease This will ensure free swinging of the buckets Most...

Page 19: ...e chemical the stress corrosion starts This type of corrosion is even more dangerous than the chemical corrosion as the effects of this corrosion are microscopic and very difficult to observe in the c...

Page 20: ...o the supply The electricity installation is not defective Power is supplied In case of below written failures related error codes are shown Err 2 Motor overheat Please switch off the centrifuge for a...

Page 21: ...20 SECTION 9 ELECTRICAL CIRCUIT DIAGRAMS 9 1 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200...

Page 22: ...21 9 2 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200R...

Page 23: ...22 SECTION 10 WARNING LABEL GROUNDED SOCKET NF 800R FUSES 2x10A NF 800 FUSES 2x4A...

Reviews: