5.4.  DRIVE SYSTEM 

  The rotor is driven by a three phase asynchronous motor. The microprocessor control system 

assures the correct drive speed. 

  The force  applied  to  the  rotor  is  directly  related  to  the  shape  of  the  rotor,  the  swing-out  rotor 

receives  more  load  than  the  angle  rotor  does.  Longer  radius  and  more  accessories  increase 
the load of the rotor and decrease the spinning speed.  

  The centrifuge does not allow the rotor to spin at a speed which it cannot resist mechanically.  

  Please make that sure the correct type of rotor is selected during programming.  

SECTION 6 

CLEANING AND PERIODIC MAINTENANCE 

6.1.  PERIODIC MAINTENANCE 

 

  Disconnect  the  power  cable  and  after  the  maintenance  check  the  presence  of  the  mains 

grounding line. 

  Rotors should be washed after every use, especially if a spillage has occurred, in warm water 

containing a few drops of liquid soap. (A mild washing liquid is ideal as a cleaner). 

  Rotors and oth

er accessorıes must be clean if any spillage or chemicals occur. 

  You may use a nylon brush to clean the buckets and tube inserts of the rotor. 

 

Do not use metal brushes. 

  Dry the rotor with a piece of soft absorber cloth. Please make sure that the buckets and inserts 

are well dried, you may use hair dryer.

 

  The buckets of the swing-out rotors should be greased frequently with the oil provided with the 

centrifuge. Please remove the light oil from the pins and put a small amount of fresh oil every 
time you grease. This will ensure free swinging of the buckets. Most of the imbalance problems 
are mostly raised by the users who do not clean and oil the pins. 

  Please do not leave the rotor on a metal surface, particularly stainless steel as electrochemical 

reactions set off easily with the aluminum or magnesium in the rotor. 

  Make sure that no deposit remains at the bottom of the bucket because the pressure of a flask 

or tube from above during centrifugation will certainly increase the chance of corrosion. 

6.2.  STERILIZATION 

  Apply alcohol, for example %70 ethanol or isoprophanol, for 10 minutes against bacteria and 

viruses. 

 

The  rotors  and  buckets  may  be  autoclaved  at  121°C  and  under  215  kPa  pressure  for  20 
minutes but please do not forget to remove all accessories. 

  Do not use formaldehyde for the sterilization process. 

  Phenol is a corrosive substance and should never be used. 

  Glutaraldehyde is a toxic substance and increases the rate of fatty acid in the body. 

6.3.  CORROSION INFORMATION 

  Nuve rotors which are made of aluminum are designed to spin at proper RCFs for many years. 

When used properly, their resistance to corrosion and their life span increases and the 
imbalance problems decrease. 

  All accessories should be checked thoroughly and regularly as almost all laboratories already 

have the conditions which lead to corrosion easily.  

Summary of Contents for NF 1200

Page 1: ...0 N VE SANAY MALZEMELER MALAT VE T CARET A NF 1200 1200 R MULTI PURPOSE BENCH TOP CENTRIFUGES USER S MANUAL Z14 K25 249 Rev No 07 Rev Date 01 2016...

Page 2: ...lity in the case where the goods are not used in accordance with their proper intent 4 The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper hand...

Page 3: ...ove the warning labels do not operate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your N ve agent for...

Page 4: ...CTING THE MEMORIZED PROGRAM 13 4 5 SAFETY INTERLOCK SYSTEM 14 4 6 IMBALANCE DETECTION SYSTEM 14 4 7 MANUAL LID OPENING 14 SECTION 5 15 OPERATING PRINCIPLES 15 5 1 PREPARATION OF THE ROTOR TO RUN 15 5...

Page 5: ...h a cooling system which provides to operate between 9 C and 40 C It ensures safe processes by means of the locking system which does not allow the centrifuge to operate if the lid is open and which d...

Page 6: ...ge 01 99 min Hold Position Timer Step 1 min Acceleration Rate 0 Slowest 9 Fastest Braking Rate 0 Slowest 9 Fastest Power supply 230 V 50 Hz Temp range 9 C 40 C Refrigerant Liquid R134a Temp Accuracy 1...

Page 7: ...daptors 1x50 ml G 51 019 Set of 6 adaptors 1x50 ml cocinal G 51 020 Set of 6 adaptors 1x30 38 ml G 51 021 Set of 6 adaptors 1x15 ml cocinal B 50 022 RS 100 Angle rotor 10x10 ml 12 000 rpm 14 167xg G 5...

Page 8: ...mm Max Speed rpm Max RCF xg RA280 Swing out Rotor 4x280ml 164 4 100 3 082 Set of 4 buckets 380 ml nominal 76 Set of 4 inserts 1x280ml 65 Set of 4 inserts 1x250ml flat 62 Set of 4 inserts 1x200ml flat...

Page 9: ...ailure The working bench should be durable to the device weight and vibration isolated Ensure that the rotor is placed correctly prior to usage According to the standard IEC 61010 2 020 anyone and any...

Page 10: ...d motor shaft IF MENTIONED WARNINGS ARE NOT CONSIDERED N VE WILL NOT BE RESPONSIBLE FROM THEIR RESULTS SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol...

Page 11: ...y of the run time from 1 min to 99 min and hold position H 03 Acceleration Brake Temperature Display Display of the acceleration and brake rates from 0 to 10 for NF 1200 R it also displays the bowl te...

Page 12: ...Keys They increase or decrease values during programming stage and to select the program number during stand by position By pushing value increase key it is possible to see the set values during centr...

Page 13: ...on the time display 2 flashes Enter the run time by pushing the value increase and decrease keys 16 H means hold position Push the function key 16 and see that the acceleration led turns on and the v...

Page 14: ...tion led turns on The set speed or RCF value is displayed during the set time duration RPM or RCF screen will appear in every 3 seconds For the operations whose speeds are higher than 2 000 rpm the ro...

Page 15: ...electronically when an unacceptable imbalance occurs In this case the brake is applied immediately The Err 1 imbalance error is displayed and the imbalance indicator turns on The lid can not be opene...

Page 16: ...es should be at the same height to balance the load Dynamic Balance This balance states that diametrically oppositely replaced center of the gravity of tubes are symmetrical with respect to spinning a...

Page 17: ...t the rotor fits on the drive shaft See Figure 4 Screw the shaft nut with socket wrench to the clockwise direction Make sure that shaft nut is screwed tightly but do not expose over force to the socke...

Page 18: ...quently with the oil provided with the centrifuge Please remove the light oil from the pins and put a small amount of fresh oil every time you grease This will ensure free swinging of the buckets Most...

Page 19: ...e chemical the stress corrosion starts This type of corrosion is even more dangerous than the chemical corrosion as the effects of this corrosion are microscopic and very difficult to observe in the c...

Page 20: ...o the supply The electricity installation is not defective Power is supplied In case of below written failures related error codes are shown Err 2 Motor overheat Please switch off the centrifuge for a...

Page 21: ...20 SECTION 9 ELECTRICAL CIRCUIT DIAGRAMS 9 1 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200...

Page 22: ...21 9 2 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200R...

Page 23: ...22 SECTION 10 WARNING LABEL GROUNDED SOCKET NF 800R FUSES 2x10A NF 800 FUSES 2x4A...

Reviews: