NUVE EN 032 Instruction Manual Download Page 8

 

5.

 

INSTALLATION 

5.1.

 

ENVIRONMENTAL CONDITIONS 

The instrument is design to operate safely under the following conditions: 

 

Indoor use only 

 

Ambient temperature: 5°C to 40°C 

 

Maximum relative humidity for temperature up to 22°C:  80% 

 

Maximum altitude: 2000 m 

 

Temperature for maximum performance:  15°C / 25°C 

 

The instrument is designed to operate in 20°C (±5°C) laboratory conditions. The 
efficiency of the unit may decrease if the ambient temperature exceeds the limits. 

5.2.

 

HANDLING AND TRANSPORTATION 

All handling and transportation must be carried out by using proper equipment and experienced 
staff. The instrument must be supported underneath and never be turned upside down. 

5.3.

 

UNPACKING 

Remove the cardboard box packing and the second nylon wrapping around the instrument. The 
below mentioned are provided with the instrument, please check them; 

 

1 ea. user’s manual and warranty 

 

1 ea. electrical connection cable  

 

2 ea. shelves  

 

4 ea. shelf carriers  

5.4.

 

MAINS SUPPLY 

The instrument requires 230V, 50/60 Hz. 
Please make sure that the supplied mains matches the required power ratings which are written 
on the name of plate of the instrument located at the back of the incubator. 

 

Always plug-in the instrument to correctly grounded sockets. 

 

 

A supply fitted with a circuit breaker should be used for protection against indirect 
contact in case of isolation fault. 

5.5.

 

POSITIONING 

 

Check that no damage has occurred during transportation. 

 

Check that the positioning is suitable for the users. 

 

Check that the instrument is stable on its four feet. If necessary, provide stable standing by 
adjusting the pedestal heights. 

 

Check  that  the  user  will  be  able  to  follow  up  the  operation  even  when  he  deals  with 
something else.  

 

Check that the positioning of the device prevents interference with other equipment in the 
near surrounding. 

Summary of Contents for EN 032

Page 1: ...N VE SANAY MALZEMELER MALAT VE T CARET A EN 032 EN 055 EN 120 INCUBATORS INSTRUCTION MANUAL Z14 K25 230 Rev No 03 Rev Date 01 2016...

Page 2: ...e damaged instrument 4 do not operate the instrument with a damaged cable 5 do not move the instrument during operation In case of a problem contact your N ve agent for an authorized service or mainte...

Page 3: ...nditions use of the product in non standard environmental conditions including but not limited to failure to meet requirements of ambient temperature lubrication humidity or magnetic field influences...

Page 4: ...UNPACKING 8 5 4 MAINS SUPPLY 8 5 5 POSITIONING 8 5 6 GENERAL PRESENTATION 9 5 7 CONTROL PANEL 10 5 8 PRIOR TO OPERATION 10 6 OPERATING PRINCIPLES 11 6 2 COMPLETION OF THE WORK 12 7 PERIODIC MAINTENANC...

Page 5: ...natural air circulation Stable and homogeneous temperature in the useful volume is provided by means of the forced air circulation through the stainless steel heaters placed onto the three outer surfa...

Page 6: ...ard maximum pcs 2 6 2 7 2 10 Power Consumption 300 W 350 W 400 W Power Supply 230 V 50 60 Hz Internal Material Stainless Steel External Material Epoxy polyester painted steel Internal Dimensions W x D...

Page 7: ...not destroy the structure of the samples Liquids are not heated in sealed containers The boiling points of the samples should be higher than the set temperature Liquids which may expand during heatin...

Page 8: ...h the instrument please check them 1 ea user s manual and warranty 1 ea electrical connection cable 2 ea shelves 4 ea shelf carriers 5 4 MAINS SUPPLY The instrument requires 230V 50 60 Hz Please make...

Page 9: ...pace between the equipment and wall 5 6 GENERAL PRESENTATION Figure 1 VENTILATION ADJUSTMENT BUTTON DISPLAY AND CONTROL PANEL DOOR HANDLE PLASTIC PADS ON OFF SWITCH FUSE HOLDER AND SUPPLY INLET VENTIL...

Page 10: ...Temperature Value Increase and Decrease Keys These keys are used to increase or decrease the values on the temperature display 7 Time Set Key This key is pushed to set the time 1 minute 99 9 hours an...

Page 11: ...key t n will appear on the temperature display Set the time value by pushing the value increase and decrease keys on the time adjustment side 1 minute to 99 9 hours Hold Push the time set key again t...

Page 12: ...maintenance 7 2 CLEANING Clean the incubator and at room temperature after disconnecting the power cable Clean the incubator with a wet cloth to remove dirt and dust Use liquid detergent to remove tou...

Page 13: ...properly The installation of the plug is not defective The mains supply is present If the incubator does not heat check the followings The program is started The safety thermostat value is adjusted hi...

Page 14: ...14 10 ELECTRICAL CIRCUIT DIAGRAM 10 1 ELECTRICAL CIRCUIT DIAGRAM FOR EN 032 10 2 ELECTRICAL CIRCUIT DIAGRAM FOR EN 055 AND EN 120 HEATERS EN 055 2 x 700 W EN 120 2 x 800 W...

Reviews: