5) The values on the displays will continue to change until they settle on the readings for the
current gas. Each of the Analyzers performs independently, and will display data in
accordance to their individual readings. User must read and consult the instructions for
each analyzer section in this manual to understand responses of each analyzer.
6)
If verifying calibration, you may now attach the calibration test gas and verify the readings are
appropriate for the test gas mix (recommended is 1000ppm CO
2
, 10ppm CO, 40% O
2
, Nitrogen
balance provided by Nuvair).
a.
If the values do not coincide with the test gas mix, refer to individual analyzer instructions
for calibration and troubleshooting.
Gas, even under moderate pressures, can cause extreme bodily harm. Never allow any
gas stream to be directed at any part of your body.
Never expose the sensor to pressures above atmospheric pressure or you may cause
damage to the sensor and/or receive false readings. Damaged Sensors will not provide
accurate gas analysis. Inaccurate gas analysis can lead to serious personal injury or
death.
It is very important that the calibration take place at atmospheric pressure and with
calibrated test gas that has no moisture in it.