Pro Trio Analyzer Manual
Page
10
NUVAIR
www.nuvair.com
3.0
Requirements for Operation of Analyzer Trio
The Analyzer Trio will require the following connections.
1.
Constant gas flow reduced to 1.0 to 5.0 L/min at ATA.
2.
Either proper electrical supply to the cord for non-battery units, or 110Volt, 60Hz AC
Outlet to power the lithium battery charger if so equipped.
3.1 Gas Flow
Before using the Analyzer Trio you must connect the analyzer to your
compressor system by diverting regulated gas to the analyzer. This will
require the use of a flow restrictor to reduce the gas to 1.0 – 5.0 L/Min
input. Nuvair sells 3 L/Min flow restrictors for this purpose. See Spares
and Accessories section.
3.2 AC Power
Use the provided AC power supply to plug into 110V power.
For lithium battery powered analyzers, the battery charger may be used as a constant power
supply. It is recommended that should the user choose to do this, the lithium battery should be
disconnected from the system to avoid damage to the cells.
3.3 Gas Sensors
The Analyzer Trio uses three individual analyzers and sensors to measure content in gases of
O
2
, CO
2
and CO. The O
2
, CO2 and CO Sensors are disposable and user-replaceable. The life
expectancy of the O2 sensor is 3 years. The life expectancy of the CO sensor is 4 years. The
life expectancy of the CO
2
sensor is 5 years. The Sensors are designed for use at atmospheric
pressure. The sample gas mixture to be analyzed must be regulated and any potential pressure
or vacuum must be avoided. Vibration, oil, moisture and particles (contamination) will shorten
the life expectancy of sensors.