Pro H2O Alarm Moisture Analyzer 

NUVAIR 

Page 

www.nuvair.com

 

 

Never expose the sensor to pressures above atmospheric pressure (0 P.S.I.) or you may 
cause damage to the sensor and/or receive false readings.  Damaged Sensors will not 
provide accurate gas analysis.  Inaccurate gas analysis can lead to serious personal 
injury or death.

 

 

 

It is very important that the calibration take place at atmospheric pressure (1 bar) and 
with calibrated test gas that has no moisture in it. 

 

 

The Pro H

2

O

TM

 can be used to monitor a regulated gas sample flow, the contents of a gas 

cylinder, or the flow from a regulator: 
 

  If analyzing a gas flow, the Sample Flow Method is the preferred method.  The flow rate 

must equal 0.5 to 1 L/min at atmospheric pressure (1 bar).  To produce this flow, a Flow 
Restrictor and Regulator may be required.  Contact Nuvair if you need assistance. 

 

4.1. Sample Flow Method of Checking Calibration (Preferred) 
 

 

 

 

 

 
 
 
 
 
 
 
 
 
 

 

Nuvair offers gas testing kits see addendum for more information.  
Testing Gas must not contain Moisture for calibration to succeed.  

 

 
 
 
 
 
 
 
 

Step 1. Attach 

Flexible Tubing 

to Gas Sample 

Flow of 0.5 to 1 

L/min

 

Step 2. Verify 

that Gas is 

Flowing Out 

Holes in Flow 

Adapter Cap

 

Step 3. Allow 30 

Seconds for Display 
Reading to Stabilize

 

Step 4. Record 

Reading while Gas is 

Flowing

 

 

 

 

Flow Restrictor/ 

Regulator 
Assembly 

Moisture Free 

Calibration Gas 

Canister 

Summary of Contents for Pro H2O Alarm

Page 1: ...Operation Manual Pro H2O AlarmTM Moisture Analyzer Rev 08 17...

Page 2: ...important safety information and should always be available to those personnel operating this equipment Read understand and retain all instructions before operating this equipment to prevent injury o...

Page 3: ...ow Method 9 4 2Programming Procedures 10 4 3Alarm Setting 11 4 4 Full Scale Value Setting 11 4 5 PrE Pressure Value 11 5 0Calibration 12 6 0Threshold Alarms 12 7 0 Powering Off 12 8 0 Factory Reset 12...

Page 4: ...ormation introduced by the following signal words Indicates an imminently hazardous situation which if not avoided will result in serious personal injury or death Indicates a potentially hazardous sit...

Page 5: ...includes a Digital Display and controls that are environmentally sealed The Analyzer uses a Flow Adapter Cap and Flexible tubing to deliver sample gas to the Sensor Pressurized gases must be regulated...

Page 6: ...le and user replaceable with a life expectancy of 60 months plus depending on usage The Sensor is designed for use at atmospheric pressure 0 P S I The gas mixture to be analyzed must be regulated acco...

Page 7: ...incorrect readings and may damage the Analyzer Incorrect readings will deliver inaccurate moisture readings Checking Calibration or use of the Analyzer with a low battery may result in inaccurate read...

Page 8: ...t key it is displayed the Temperature value C TeM 4 1 Pressing again the ADJUST key PPM in volume PPM is displayed 4 2 Pressing again the ADJUST key Absolute humidity gr m3 AbH is displayed The absolu...

Page 9: ...he flow from a regulator If analyzing a gas flow the Sample Flow Method is the preferred method The flow rate must equal 0 5 to 1 L min at atmospheric pressure 1 bar To produce this flow a Flow Restri...

Page 10: ...ion o AL 2 Second alarm point expressed in ppm of Moisture in concentration o FSC Value expressed in ppm of moisture in concentration corresponding to the current output full scale value 4 20mA output...

Page 11: ...button On the display will appear Pr for two seconds then AL 1 will appear Press the On Off button to cycle through the Alarm settings until you reach the FSC view The FSC screen and a 3 digit value w...

Page 12: ...ading Press and hold the On Off and Prog buttons simultaneously until the display flashes ALN then ON The alarm function is active when a triangle is present in the upper left of the analyzer display...

Page 13: ...is may result in personal injury or death Do not clean Analyzer with anything other than a damp soft cloth Do not immerse in liquid leave unprotected outside or store in a wet environment Protect Anal...

Page 14: ...livered at a specific flow rate and pressure Nuvair offers a calibration gas canister with a compatible flow restrictor and regulator assembly to regulate the gas for proper calibration 11 0 Troublesh...

Page 15: ...ecial Mini Jack plug that can be purchased separately from Nuvair Insert the plug into the output jack Connections diagram below for additional wiring The plug should be connected or disconnected when...

Page 16: ...han 10 C and greater than 40 C additional error rh 0 25 rh rh for less than 10 rh and greater than 90 rh Relative Humidity Sensor Type Capacitive Expected Sensor Life Room Air 5 Years Power 1 9 Volt B...

Page 17: ...l begin with the date of manufacture by NUVAIR Maintenance Items Any materials which are consumed or otherwise rendered not warrantable due to processes applied to them are considered expendable and a...

Page 18: ...Pro H2O Alarm Moisture Analyzer NUVAIR Page 18 www nuvair com Notes...

Page 19: ...Pro H2O Alarm Moisture Analyzer NUVAIR Page 19 www nuvair com...

Page 20: ...Nuvair Phone 805 815 4044 Fax 805 486 09100 1600 Beacon Place Oxnard CA 93036 USA Email info nuvair com Web www nuvair com Revision 08 14...

Reviews: