Pro H2O Alarm Moisture Analyzer 

NUVAIR 

Page 

www.nuvair.com

 

2.6. Flow Adapter Cap 

 
The Analyzer includes a Flow Adapter Cap with flexible tubing and flow orifice.  It attaches to the 
Sensor port and is sealed by an o-ring.  It can be used to direct the gas sample flow to the 
Sensor via one of two methods: 
 
 
 
 
 
 
 
 
 
 
Flow to the sensor needs to be restricted to .5-1liter of pressure maximum. Nuvair offers flow 
restrictors to accomplish this task. See appendix. 
 

3.0 Calibration 

 

Analyzer calibration must be verified on a daily basis.  Improper calibration may result in 
an incorrect reading, delivering inaccurate levels of moisture being monitored.  

 

 

This Analyzer must always be checked with a calibration gas and used with gases 
regulated and supplied at atmospheric pressure (0 P.S.I.).  Use of gases at higher 
pressures may result in incorrect readings and may damage the Analyzer.  Incorrect 
readings will deliver inaccurate moisture readings.

 

 

 

Checking Calibration or use of the Analyzer with a low battery may result in inaccurate 
readings.  Inaccurate gas analysis can lead to serious personal injury or death. 

 

If the Analyzer has been subjected to a recent change in ambient temperature, allow it to 
stabilize for one hour before checking calibration. 
 

Verify calibration on a daily basis.  Breathing gas applications require the use of a certified 
calibration gas with no moisture concentration and flow rate of 0.5-1 L/min.  The equipment to 
produce this flow is available from Nuvair.  See Spares and Accessories section. 
 
To assure the greatest accuracy for other applications, use the calibration gas concentration 
closest to the expected concentration in the gas being measured.   
 
 

Sample Flow Method 

Flow is Delivered from 

Low Pressure Source 

via Flexible Tubing

 

Direct Flow Method 

Flow is Directed to 

Flow Orifice on Face 
of Flow Adapter Cap

 

 

Vent Hole 

Summary of Contents for Pro H2O Alarm

Page 1: ...Operation Manual Pro H2O AlarmTM Moisture Analyzer Rev 08 17...

Page 2: ...important safety information and should always be available to those personnel operating this equipment Read understand and retain all instructions before operating this equipment to prevent injury o...

Page 3: ...ow Method 9 4 2Programming Procedures 10 4 3Alarm Setting 11 4 4 Full Scale Value Setting 11 4 5 PrE Pressure Value 11 5 0Calibration 12 6 0Threshold Alarms 12 7 0 Powering Off 12 8 0 Factory Reset 12...

Page 4: ...ormation introduced by the following signal words Indicates an imminently hazardous situation which if not avoided will result in serious personal injury or death Indicates a potentially hazardous sit...

Page 5: ...includes a Digital Display and controls that are environmentally sealed The Analyzer uses a Flow Adapter Cap and Flexible tubing to deliver sample gas to the Sensor Pressurized gases must be regulated...

Page 6: ...le and user replaceable with a life expectancy of 60 months plus depending on usage The Sensor is designed for use at atmospheric pressure 0 P S I The gas mixture to be analyzed must be regulated acco...

Page 7: ...incorrect readings and may damage the Analyzer Incorrect readings will deliver inaccurate moisture readings Checking Calibration or use of the Analyzer with a low battery may result in inaccurate read...

Page 8: ...t key it is displayed the Temperature value C TeM 4 1 Pressing again the ADJUST key PPM in volume PPM is displayed 4 2 Pressing again the ADJUST key Absolute humidity gr m3 AbH is displayed The absolu...

Page 9: ...he flow from a regulator If analyzing a gas flow the Sample Flow Method is the preferred method The flow rate must equal 0 5 to 1 L min at atmospheric pressure 1 bar To produce this flow a Flow Restri...

Page 10: ...ion o AL 2 Second alarm point expressed in ppm of Moisture in concentration o FSC Value expressed in ppm of moisture in concentration corresponding to the current output full scale value 4 20mA output...

Page 11: ...button On the display will appear Pr for two seconds then AL 1 will appear Press the On Off button to cycle through the Alarm settings until you reach the FSC view The FSC screen and a 3 digit value w...

Page 12: ...ading Press and hold the On Off and Prog buttons simultaneously until the display flashes ALN then ON The alarm function is active when a triangle is present in the upper left of the analyzer display...

Page 13: ...is may result in personal injury or death Do not clean Analyzer with anything other than a damp soft cloth Do not immerse in liquid leave unprotected outside or store in a wet environment Protect Anal...

Page 14: ...livered at a specific flow rate and pressure Nuvair offers a calibration gas canister with a compatible flow restrictor and regulator assembly to regulate the gas for proper calibration 11 0 Troublesh...

Page 15: ...ecial Mini Jack plug that can be purchased separately from Nuvair Insert the plug into the output jack Connections diagram below for additional wiring The plug should be connected or disconnected when...

Page 16: ...han 10 C and greater than 40 C additional error rh 0 25 rh rh for less than 10 rh and greater than 90 rh Relative Humidity Sensor Type Capacitive Expected Sensor Life Room Air 5 Years Power 1 9 Volt B...

Page 17: ...l begin with the date of manufacture by NUVAIR Maintenance Items Any materials which are consumed or otherwise rendered not warrantable due to processes applied to them are considered expendable and a...

Page 18: ...Pro H2O Alarm Moisture Analyzer NUVAIR Page 18 www nuvair com Notes...

Page 19: ...Pro H2O Alarm Moisture Analyzer NUVAIR Page 19 www nuvair com...

Page 20: ...Nuvair Phone 805 815 4044 Fax 805 486 09100 1600 Beacon Place Oxnard CA 93036 USA Email info nuvair com Web www nuvair com Revision 08 14...

Reviews: