OM0291
Page 30 of 56
Rev 3 November2020
6.6.3 Assembly
a. Replace front grill (if removed).
b. Replace the work tray and carefully tighten the thumbscrews.
c. Replace perforated metal diffuser screen over the underside of the supply HEPA filter.
d. Wipe down all exposed surfaces of the work area with 70% isopropyl alcohol.
e. Prepare for aseptic operation.
¹ Available from Lab Safety Supply, Janesville, WI 53547-1368, or other laboratory, industrial, or hospital supply distributors.
²American Society of Hospital Pharmacists. 2006. ASHP Guidelines on Handling Hazardous Drugs
Am. J. Hosp. Pharm.
63:1172-1193.
7.0
General Maintenance
All maintenance actions on this equipment must be performed by a qualified
technician who is familiar with the proper maintenance procedures required
including both certification and repair.
7.1 Decontamination
No maintenance should be performed on the interior of the LabGard® ES cabinet (area behind access panels) unless the
cabinet has been microbiologically decontaminated, is known to be biologically clean, or known to be chemically inert.
Surface disinfection is performed as specified in the Cleaning Procedures section.
Hazardous Gases! Personal Protection Equipment Required.
Decontamination using hazardous gas, vapor or mist must be performed in accordance with the
specifications of NSF/ANSI 49, Annex G, EN12469, Annex J or applicable national, state, province
or local regulations.
This procedure presents considerable risks and must be performed only by specially trained and authorized service
personnel in accordance with applicable national safety regulations. (i.e. US – OSHA/NIOSH, Germany TRGS-522).
The decontaminate is generated either external or internal of the sealed cabinet. The quantity of decontaminate should
follow standard or manufacturer’s recommendations based on cabinet volume. The decontaminate process should follow
standards or manufacturer’s recommendations based on the decontaminate used.
All decontaminate materials are hazardous (chemical-liquid, gas and vapor)
(Flammable – process) and are required to be handled properly. Follow all
product and process documentation and labelling.
CAUTION
!
CAUTION
!
CAUTION
!
Summary of Contents for LabGard NU-640-400
Page 7: ...OM0291 Page 7 of 56 Rev 3 November2020...
Page 9: ...OM0291 Page 9 of 56 Rev 3 November2020...
Page 10: ...OM0291 Page 10 of 56 Rev 3 November2020...
Page 11: ...OM0291 Page 11 of 56 Rev 3 November2020...
Page 15: ...OM0291 Page 15 of 56 Rev 3 November2020...
Page 20: ...OM0291 Page 20 of 56 Rev 3 November2020...