LabGard® ES Energy Saver
Class II, Type A2 Animal Handling
Biosafety Cabinet
Models
NU-640-400/500/600
Manufactured by:
NuAire, Inc. - Plymouth, Minnesota, U.S.A.
1.0 General Information
1.1 Description
The LabGard® ES Model NU-640 Animal Handling Biosafety Cabinet (AHBSC) is a bench/table top model, optionally available
with a base support stand, for operation as a console model. The LabGard® ES model NU-640 utilizes an Energy Saver DC
ECM motor optimally determined forward curved fan for each model size/width to maximize both energy efficiency and
filter loading capacity. The Energy Saver ECM motor is controlled by the Aeromax™ Control System to nominal airflow
setpoints via an integrated solid-state DC motor controller based on feedback algorithm to provide automatic
compensation (constant volume control) for both filter loading and line voltage variances. Airflow velocity is monitored by
the PresurFlow™ system with a digital pressure sensor.
The Animal Handling Biosafety Cabinet, (AHBSC) is a product resulting from the development of the "laminar flow" principle
(see Airflow Schematic) and the application of environmental controls as required in the field of biological research or
chemical containment. The AHBSC, when used with proper technique, is an effective primary engineering control in
obtaining the optimum product quality while reducing the potential for exposure of both product and personnel to airborne
biological or particulate chemical agents in low to moderate risk-hazard research and drug preparation or product
operations, as prescribed by the Center for Disease Control (CDC) Atlanta, Georgia.
The NU-640 bench AHBSC meets the requirements of a Class II, since the cabinet conforms to the following requirements:
1. Maintains a minimum inflow velocity of 100 LFPM (.51mps) through the work access opening.
2. Has HEPA filtered downflow air that is mixed with the inflow air from a common exhaust plenum.
3. Discharges a percentage of air to the outside atmosphere after HEPA filtration.
4. Has all biologically contaminated ducts and plenums under negative pressure or surrounded by negative pressure.
Cabinets used for work with minute quantities of non-flammable or explosive volatile toxic chemicals and tracer amounts of
radionuclides required as an adjunct to microbiological studies must be exhausted through properly functioning exhaust
canopies.
Summary of Contents for LabGard NU-640-400
Page 7: ...OM0291 Page 7 of 56 Rev 3 November2020...
Page 9: ...OM0291 Page 9 of 56 Rev 3 November2020...
Page 10: ...OM0291 Page 10 of 56 Rev 3 November2020...
Page 11: ...OM0291 Page 11 of 56 Rev 3 November2020...
Page 15: ...OM0291 Page 15 of 56 Rev 3 November2020...
Page 20: ...OM0291 Page 20 of 56 Rev 3 November2020...