OM0228
Page
88
of
99
February/2014
9.0
Remote
Contacts
The
NU
‐
430
has
several
contact
closures
for
remote
sensing
of
various
functions.
9.1
Fan
Relay
The
fan
relay
contacts
are
dual
normally
open
contact
closure
outputs
which
are
activated
whenever
the
blower
is
turned
on.
Contact
ratings
are
250
VAC
maximum
at
2
Amps.
9.2
Alarm
Relay
The
alarm
relay
contacts
are
dual
normally
open
contact
closure
outputs
which
are
activated
whenever
an
airflow
alarm
condition
occurs.
An
airflow
alarm
condition
will
occur
if
either
airflow
sensor
detects
5
consecutive
2
second
airflow
readings
above
or
below
the
alarm
limits.
Contact
ratings
are
250
VAC
maximum
at
2
Amps.
9.3
AUX
Relay
The
AUX
relay
contacts
are
common,
normally
open
and
normally
closed
contact
closure
outputs
which
are
activated
whenever
the
blower
is
turned
on.
However,
the
AUX
relay
does
have
some
conditional
logic
programmed
to
aid
the
supply
On/Off
and
exhaust
delay
options.
The
relay
will
activate
whenever
the
blower
is
turned
on
and
stay
on
unless
after
5
minutes
there
is
a
low
exhaust
alarm,
then
the
relay
will
de
‐
activate.
If
exhaust
is
sufficient,
the
relay
will
stay
active.
If
the
blower
is
then
turned
off,
the
relay
will
stay
active
for
one
minute,
then
de
‐
activate.
The
AUX
relay
may
also
be
selected
to
operate
the
same
as
the
fan
relay,
reference
the
AUX
relay
function
in
the
option
menu.
Contact
ratings
are
250
VAC
maximum
at
2
Amps.
9.4
Remote
Override
The
remote
override
contacts
are
(no
power)
shorting
contacts
only,
which
when
closed,
indicates
to
the
control
system
to
shut
down
the
cabinet.
The
display
will
indicate
remote
override,
inhibit
all
alarms,
fluorescent
light,
internal
blower
and
close
the
exhaust
valve
(i.e.
NU
‐
951)
if
present.
9.5
Night
Setback
The
night
setback
contacts
are
(no
power)
shorting
contacts
only,
which
when
closed,
indicates
to
the
control
system
to
place
the
cabinet
into
night
setback.
The
display
will
indicate
night
setback
active,
inhibit
exhaust
alarms,
fluorescent
light
and
internal
blower.
Summary of Contents for Labgard NU-430-400
Page 7: ...OM0228 Page 7 of 99 February 2014 ...
Page 9: ...OM0228 Page 9 of 99 February 2014 ...
Page 10: ...OM0228 Page 10 of 99 February 2014 ...
Page 16: ...OM0228 Page 16 of 99 February 2014 ...
Page 17: ...OM0228 Page 17 of 99 February 2014 ...
Page 41: ...OM0228 Page 41 of 99 February 2014 ...
Page 93: ...OM0228 Page 93 of 99 February 2014 ...
Page 94: ...OM0228 Page 94 of 99 February 2014 ...
Page 95: ...OM0228 Page 95 of 99 February 2014 ...
Page 96: ...OM0228 Page 96 of 99 February 2014 ...
Page 97: ...OM0228 Page 97 of 99 February 2014 ...