OM0228
Page
23
of
99
February/2014
6.1.4
Standby/Run
Mode
Alarms
If
present,
standby/run
mode
alarms
will
be
both
visual
and
audible,
the
Red
LED
oval
under
the
TOUCHLINK
LCD
display
will
turn
on,
and
the
TOUCHLINK
LCD
screen
will
also
display
a
description
of
the
alarm
in
place
of
the
NuAire
Logo
along
with
alarm
silence
icon.
Depending
upon
the
alarm
type,
the
LFBSC
profile
will
also
indicate
in
red
the
alarm
present.
Audible
alarms
can
be
silenced
or
will
produce
an
alarm
tone
for
30
seconds,
then
into
a
ring
back
cycle
of
once
every
10
seconds.
Pressing
the
alarm
silence
icon
will
silence
the
audible
alarm
for
15
minutes,
then
into
a
ring
back
cycle
at
of
once
every
10
seconds.
Alarm
Types
Window
High
‐
window
is
raised
above
its
nominal
height
Downflow
High
Limit
‐
downflow
is
above
the
high
alarm
setpoint
Downflow
Low
Limit
‐
downflow
is
below
the
low
alarm
setpoint
Inflow
High
Limit
‐
inflow
is
above
the
high
alarm
setpoint
Inflow
Low
Limit
‐
inflow
is
below
the
low
alarm
setpoint
MENU
xx:xx
AM
xx/xx/xx
0.0 fpm
0.0 fpm
UV
FL
ACTIVE ALARMS!
WINDOW HIGH!
Summary of Contents for Labgard NU-430-400
Page 7: ...OM0228 Page 7 of 99 February 2014 ...
Page 9: ...OM0228 Page 9 of 99 February 2014 ...
Page 10: ...OM0228 Page 10 of 99 February 2014 ...
Page 16: ...OM0228 Page 16 of 99 February 2014 ...
Page 17: ...OM0228 Page 17 of 99 February 2014 ...
Page 41: ...OM0228 Page 41 of 99 February 2014 ...
Page 93: ...OM0228 Page 93 of 99 February 2014 ...
Page 94: ...OM0228 Page 94 of 99 February 2014 ...
Page 95: ...OM0228 Page 95 of 99 February 2014 ...
Page 96: ...OM0228 Page 96 of 99 February 2014 ...
Page 97: ...OM0228 Page 97 of 99 February 2014 ...