OM0228
Page
74
of
99
February/2014
Run
Times
This
parameter
allows
the
service
technician
to
view,
alter,
or
reset
both
UV
light
and
motor
blower
run
timer.
Select
desired
run
time
parameter
from
menu.
UV
Run
Time.
Motor
Blower
Run
Time.
UV
FL
UV
FL
MENU
xx:xx
AM
xx/xx/xx
UV
FL
UP
DOWN
ENTER
UV RUN TIME
BLOWER MOTOR RUN TIME
MENU
xx:xx
AM
xx/xx/xx
UP
DOWN
SAVE
MOTOR BLOWER RUN TIME
CURRENT HOURS = XXXX
THE MOTOR BLOWER RUN TIME INDICATES HOW
MANY HOURS THE BLOWER HAS BEEN ON. TIME
MAY BE RESET OR TRANSFERRED FOR SERVICE
PURPOSES.
PRESS UP, DOWN TO CHANGE THE HOURS.
MENU
xx:xx
AM
xx/xx/xx
RESET
UV LIGHT USAGE
CURRENT HOURS = XXXX
THIS TIME SHOWS HOW MANY HOURS THE UV
LIGHT HAS LOGGED ON. RESET THE TIME WHEN
REPLACING THE UV BULB.
PRESS MENU TO EXIT WITHOUT RESET