Ntron OxyTX
Operation and Installation Manual
Revised:10/2013 ECN00008
Part No.:01-507
Ntron Ltd, Mullaghboy Industrial Park, Navan, Co Meath, IRELAND
Tel: + 353 (46)9071333 • Fax: + 353 (46)9071331 •
Email: [email protected] • Website: www.ntron.com/ntron
Ntron Ltd, Dallam Court, Dallam Lane, Warrington WA2 7LT, UNIITED KINGDOM
Tel: + 44 (161) 930 8690 • Fax: + 44 (161) 930 8691
T
ECHNICAL
S
PECIFICATIONS
OxyTX:
Transmitter Enclosure:
ABS with Stainless Steel Identification label (IP66)
Weight:
800g
Power Requirements:
7 - 30 VDC 2 wire loop power
Output:
4 - 20 mA
Range:
0 - 25 mA
Display:
4 digit LCD display
Resolution:
0.1%
Oxygen
Operating Humidity:
0 - 95% RH, non-condensing
Operating Temperature:
-20°C to 45°C
Approvals:
ATEX, Ex II 2 G, EEx ia IIC T4
Oxygen Sensor - Sample:
Flow:
100 ml/min minimum
Pressure:
4
PSIG
maximum
Temperature:
-5°C to 50°C
Oxygen Sensor:
Type:
Electrochemical fuel cell
Sensor life:
Typically
12
months
Operating Humidity:
15-90%
0-99%
intermittent
Calibration Adjustments:
Zero and span, located on front panel.
Certification:
ATEX, Ex II 1 G, EEx ia IIC T6
P
AGE
16
OF
19