Setting 4mA Level (Menu Mode C:4):
This calibration feature allows the instrument to simulate a condition of zero gas so
that the 4mA output can be set.
•
Attach a multimeter (set to measure DC voltage) between test points TP1 and
TP2.
•
Select menu mode
C: 4
and press ENTER.
•
Using the UP and DOWN buttons, adjust the reading displayed on the
multimeter to 40mV (+/- 0.5mV)
•
Press ENTER to store the 4mA calibration data. Pressing MENU instead of
ENTER aborts this feature.
•
Press MENU – the display will show the DAC 4mA calibration factor for the
instrument before returning to its standard mode of operation.
Setting 20mA Level (Menu Mode C:5):
This calibration feature allows the instrument to simulate a condition of full-scale gas
so that the 20mA output can be set. A control unit connected will indicate full-scale
gas also and may enter its alarm state.
•
Attach a multimeter (set to measure DC voltage) between test points TP1 and
TP2.
•
Select menu mode
C: 5
and press ENTER.
•
Using the UP and DOWN buttons, adjust the reading displayed on the
multimeter to 200mV .
•
Press ENTER to store the 4mA calibration data. Pressing MENU instead of
ENTER aborts this feature.
Display Mode (Menu Mode C:6):
•
Select menu mode
C: 6
and press ENTER.
•
Use the UP and DOWN buttons to move the decimal point to the desired
setting.
•
Press ENTER. Pressing MENU instead of ENTER aborts the feature.
•
Press MENU to return the instrument to its standard mode of operation.
Note: Changing the decimal point will not result in a more accurate reading.