View Engineer/Diagnostics Data (Menu Mode C:9):
This feature is a
view-only
feature. No configuration changes are possible from
within this menu.
This information assists in trouble shooting fault conditions present on the OxyTx.
•
Select menu mode
C: 9
(refer to section 3) and press ENTER.
•
The display will alternate between the current value and code
C: 9x
: Where is
x is: 0
Sensor reading.
1
Calibration temperature,
8
C.
2
Temperature compensation value
8
C.
3
Sensor temperature,
8
C.
4
Detector AtoD counts.
5
Reference AtoD counts.
•
The mode of operation can be selected by pressing the UP button.
•
Press MENU to return the instrument to its standard mode of operation.
Cross Reference (Menu Mode C:12):
No customer adjustment of this value is necessary. The set value of “
1.00”
should not
be changed.
Positive Zero Suppression (Menu Mode C:19):
No customer adjustment of this value is necessary. The set value of “
0”
should not be
changed.
Negative Zero Suppression (Menu Mode C:20):
No customer adjustment of this value is necessary. The set value of “
0”
should not be
changed.
(Menu Mode C:30):
No customer adjustment of this value is necessary. The set value of “
OFF”
should
not be changed.