Ntron OxyTX
Operation and Installation Manual
Revised:10/2013 ECN00008
Part No.:01-507
Ntron Ltd, Mullaghboy Industrial Park, Navan, Co Meath, IRELAND
Tel: + 353 (46)9071333 • Fax: + 353 (46)9071331 •
Email: [email protected] • Website: www.ntron.com/ntron
Ntron Ltd, Dallam Court, Dallam Lane, Warrington WA2 7LT, UNIITED KINGDOM
Tel: + 44 (161) 930 8690 • Fax: + 44 (161) 930 8691
OxyTx Head Indications
Normal Conditions
The Oxytx Transmitter Head will display a steady reading for gas levels between 0
and FSD.(Full Scale Deflection)
High Gas Conditions (System Fault Conditions)
The OxyTx display will flash between the gas reading and ‘Hi’ when the gas reading
is above the sensor FSD. This will coincide with the head drawing a current of less
than 2.5mA from the control unit, thus ensuring the control unit is aware of the fault
condition.
Note:
xx is a number that defines a particular fault as follows:
1
Checksum error.
2
Zero calibration error.
4
Span calibration error.
8
Sensor reference low output.
16
Sensor reference high output.
32
Sensor detector low output.
64
Sensor detector low output.
P
AGE
14
OF
19