TEC-B-01 User Manual
____________________________________________________________________________________________________________________________
__________________________________________________________________________________
version 1.1
page 12
Figure 5: schematic diagram of the TEC passive cell model. For TEC-B-01, there is no REF connector
and R
bath
= 0
Ω
.
4.2. Basic
settings
Before using the TEC-B-01 always make the basic settings to avoid oscillations.
Basic settings
Turn all controls to low values (less than 1) and each symmetrical offset adjustment, i.e. C.
HEADSTAGE BIAS CURRENT, CURRENT ELECTRODE OFFSET and POTENTIAL
OFFSET potentiometers, in the range of 5 (zero position, see chapter 2.2). Set the CAPACITY
COMPENSATION (#
13
) to OFF.
Disable the HOLD unit by setting the + / 0 / - switch (#
7
) to 0.
Set the MODE OF OPERATION (#
1
) to CC.
Set the display to POTENTIAL ELECTRODE using switch #
10
.
Now the TEC-B-01 is ready for an initial check with the cell model.