3.2. Description of the Front Panel 

 
(

1

)  POWER LED 

 
LED indicating that the ISO-STIM 01D is powered on 
 

 
(

2

)  BIPOLAR ON/ BIPOLAR OFF switch 

 

Switch that selects the stimulus mode. Switching to BIPOLAR ON results in a stimulus 
with the same amplitude in positive and negative direction without any delay. Amplitude 
and duration settings are dependent on the mode of operation and/or the settings at the 
front panel or at the external stimulus generator 
BIPOLAR stimulus mode is only functional in GATE TTL INPUT MODE 

 
(

3

)  DURATION potentiometer 

 

Potentiometer that sets the duration of the stimulus (multiplied with the duration range 
factor) generated by the built-in timing unit (ANALOG INPUT MODE, GATE TTL 
MODE). The duration range factor is set by switch 

#4

 

 

(

4

)  DURATION range switch 

 

Switch for setting the DURATION range factor (x10 µs, x 100µs, x1 ms) 
 

 

Note

: Setting of 00 as duration is 

not

 defined and should 

not

 be set!! 

 
(

5

)  AMPLITUDE x0.1V / x CURRENT RANGE potentiometer 

 

Potentiometer that sets the amplitude of the stimulus generated by the internal 
timing  unit  (GATE  TTL  MODE).  In  VOLTAGE  output  mode  the  setting  is 
multiplied by 0.1 V. In CURRENT output mode the setting is multiplied by the 
setting of the CURRENT RANGE switch (

#6

). Switch 

#7

 sets the polarity of the 

stimulus 

 
(

6

)  CURRENT RANGE switch 

 
Switch for setting the CURRENT RANGE factor (x0.1 µA, x1 µA, x10 µA) 
 
 

(

7

)  +/0/- switch 

 
Switch  for  setting  the  polarity  of  the  stimulus.  In  “0”  position  the  internal  stimulus 
generation is disabled 

 

Summary of Contents for ISO-STIM 01D

Page 1: ...FOR THE ISO STIM 01D STIMULUS ISOLATION UNIT 100 V 10 mA bipolar output VERSION 4 3 npi 2016 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 suppor...

Page 2: ...nts 4 3 ISO STIM 01D Stimulus Isolation Unit 5 3 1 System Description 5 3 2 Description of the Front Panel 6 4 Operation 9 4 1 Connections Noise Reduction 9 4 2 FILTERED UNFILTERED switch rear panel 9...

Page 3: ...re connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact connected to ground protective earth Before opening the cabinet unplug the instrume...

Page 4: ...ON Observe extreme caution when working with this instrument 1 Always connect high voltage power supplies to protective earth 2 Do not touch connections unless the instrument is turned off and the cap...

Page 5: ...mplitude and polarity The ISO STIM 01D has three modes of operation In the DIRECT input mode the output signal follows exactly the input signal In the GATE TTL input mode the output signal is generate...

Page 6: ...URATION potentiometer Potentiometer that sets the duration of the stimulus multiplied with the duration range factor generated by the built in timing unit ANALOG INPUT MODE GATE TTL MODE The duration...

Page 7: ...OUTPUT MODE switch set to CURRENT The polarity of the output signal is set by switch 7 Important If the CAP COMP 15 is overcompensated in CURRENT MODE the stimulus isolator will ring and not work pro...

Page 8: ...VOLTAGE trim pot Trim pot to compensate for the VOLTAGE OFFSET of the stimulating electrode in TTL mode It is recommended to compensate the offsets only in a completely warmed up condition i e after...

Page 9: ...possible source of noise is the chassis of the stimulator In a noisy environment it may act as an antenna picking up 50 Hz or 60 Hz noise respectively especially when operating with current stimuli In...

Page 10: ...DE VOLTAGE i e an input of 1 V will lead to an output stimulus of 10 V In CURRENT OUTPUT MODE the amplitude of the output stimulus is again dependent on the setting the CURRENT RANGE switch With the C...

Page 11: ...voltage divider between the electrode resistance and the protection resistor 4 4 ISO CAB optional accessory npi provides an optional shielded cable for connection of stimulation electrodes One end is...

Page 12: ...__________________________________________________________________________________________________ ___________________________________________________________________________ version 4 3 page 12 Figur...

Page 13: ...osition CURRENT CURRENT RANGE switch set to x1 A and the amplitude of the input signal is 2 V The max signal at the isolated output plugs has to be 20 V according to the current of 200 A flowing throu...

Page 14: ...with 3 position rotary switch 100 V voltage source Time range factor x10 s x100 s x1 ms selectable with toggle switch time range up to 99 ms Amplitude potentiometer 3 digits 10 V amplified internally...

Reviews: