ISO-STIM 01D User Manual
_________________________________________________________________________________________________________________
___________________________________________________________________________
version 4.3
page 14
6. Technical Data
Input (operation) modes:
DIRECT, ANALOG TRIGGER, TTL TRIGGER
selectable with toggle switch
Output modes:
voltage source, current source
selectable with toggle switch
Source ranges:
±100 µA, ±1 mA, ±10 mA (current source), selectable with 3-
position rotary switch
±100 V (voltage source)
Time range factor:
x10 µs, x100 µs, x1 ms
selectable with toggle switch, time range up to 99 ms
Amplitude potentiometer:
3 digits, ±10 V, amplified internally with gain of 10 (voltage
source) or scaled with source range switch (current source);
polarity selectable with toggle switch
Time potentiometer:
2 digits, 0 – 99 as multiplication factor selectable
Input voltage range:
±10 V, amplified internally with gain of 10
Input impedance:
10 k
Output:
monopolar or bipolar, selected by toggle switch
Output voltage:
±100 V max., isolated
Output current:
±10 mA max., isolated
Output Impedance:
1 k
(short circuit protection)
Dimensions:
245 x 260 x 90 mm³