ISO-STIM 01D User Manual 

_________________________________________________________________________________________________________________

 

___________________________________________________________________________ 
version 4.3 

 

 

page 13 

5.  Test Procedures 

 
For the following test procedures a function generator is required. The frequency for testing 
should be around 100 Hz, waveform is triangle. The value of the test pulse amplitude is always 
described in detail for each of the following steps. In this test condition a load resistor of 100 k

 

is connected to the output plugs (red and black plugs).  

5.1. DIRECT  INPUT  MODE 

Settings

: INPUT MODE switch is in position DIRECT, OUTPUT MODE switch in position 

VOLTAGE.  Analog  input  voltage  from  a  function  generator  is  ±2 V  and  connected  to  the 
SIGNAL INPUT BNC. Max. output voltage has to be ±20 V. 
 

Caution

If  you  measure  the  output  signal  with  an  oscilloscope  take  care  that  the  function 

generator  has  not  the  same  ground  connection  because  of  the  isolation  of  the  output.  It  is 
important to realize that a connection is also possible through the protective ground connections 
between  oscilloscope  and  function  generator  (most  mains  supplied  oscilloscopes  have  a 
connection between ground and protective earth)! Thus, we recommend to use a battery driven 
oscilloscope for test measurements to avoid unwanted grounding!

 

 
Now the OUTPUT MODE switch is set in position CURRENT (CURRENT RANGE switch 
set to x1 µA) and the amplitude of the input signal is ±2 V. The max. signal at the isolated 
output plugs has to be ±20 V according to the current of ±200 µA flowing through the load 
resistance of 100 k



 

5.2. GATE TTL INPUT MODE 

Settings

: INPUT MODE switch is in position GATE TTL, OUTPUT MODE switch in position 

VOLTAGE. The TTL output of the function generator has to be connected to the GATE TTL 
BNC, the SIGNAL INPUT BNC is not connected. The output voltage will be preset from the 
value  dialed  at  the  AMPLITUDE  potentiometer  and  the  polarity  from  the  position  of  the 
corresponding  toggle  switch.  Remember  at  any  time  Ohm’s  law  to  realize  what  potentials 
should be generated theoretically; the maximum output voltage of the system is ±100V. 

5.3. ANALOG  INPUT  MODE 

Settings

: INPUT MODE switch is in position ANALOG TRIG, OUTPUT MODE switch in 

position  VOLTAGE.  Toggle  switch  for  DURATION  range  in  position  100 µs,  TIME 
potentiometer  dialed  on  value  3.  Input  voltage from  a  function  generator  has  to  be  positive 
referring to ground, value 5 V and connected to the ANALOG INPUT BNC. Output voltage 
will occur if the input voltage is higher than 300 mV (this is the internal analog trigger level) 
and only for the desired time (here 300 µs). The output signal can be varied by using other input 
amplitude values and / or several time settings. 
 

Caution

If  you  are  working  in  CURRENT  MODE  realize  that  the  input  resistance  of  an 

oscilloscope is normally 1 M

, e.g. an oscilloscope connected to the outputs is a 1 M

 load 

for the current source producing an output signal that is measured.

 

 

Summary of Contents for ISO-STIM 01D

Page 1: ...FOR THE ISO STIM 01D STIMULUS ISOLATION UNIT 100 V 10 mA bipolar output VERSION 4 3 npi 2016 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 suppor...

Page 2: ...nts 4 3 ISO STIM 01D Stimulus Isolation Unit 5 3 1 System Description 5 3 2 Description of the Front Panel 6 4 Operation 9 4 1 Connections Noise Reduction 9 4 2 FILTERED UNFILTERED switch rear panel 9...

Page 3: ...re connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact connected to ground protective earth Before opening the cabinet unplug the instrume...

Page 4: ...ON Observe extreme caution when working with this instrument 1 Always connect high voltage power supplies to protective earth 2 Do not touch connections unless the instrument is turned off and the cap...

Page 5: ...mplitude and polarity The ISO STIM 01D has three modes of operation In the DIRECT input mode the output signal follows exactly the input signal In the GATE TTL input mode the output signal is generate...

Page 6: ...URATION potentiometer Potentiometer that sets the duration of the stimulus multiplied with the duration range factor generated by the built in timing unit ANALOG INPUT MODE GATE TTL MODE The duration...

Page 7: ...OUTPUT MODE switch set to CURRENT The polarity of the output signal is set by switch 7 Important If the CAP COMP 15 is overcompensated in CURRENT MODE the stimulus isolator will ring and not work pro...

Page 8: ...VOLTAGE trim pot Trim pot to compensate for the VOLTAGE OFFSET of the stimulating electrode in TTL mode It is recommended to compensate the offsets only in a completely warmed up condition i e after...

Page 9: ...possible source of noise is the chassis of the stimulator In a noisy environment it may act as an antenna picking up 50 Hz or 60 Hz noise respectively especially when operating with current stimuli In...

Page 10: ...DE VOLTAGE i e an input of 1 V will lead to an output stimulus of 10 V In CURRENT OUTPUT MODE the amplitude of the output stimulus is again dependent on the setting the CURRENT RANGE switch With the C...

Page 11: ...voltage divider between the electrode resistance and the protection resistor 4 4 ISO CAB optional accessory npi provides an optional shielded cable for connection of stimulation electrodes One end is...