background image

EXT-02F User Manual 

___________________________________________________________________________ 
version 4.2 

 

 

page 9 

(3)

 

GROUND connector 

Banana  plug  providing  ground.  The  plug  is  connected  internally  to  the  shields  of  the  BNC 
connectors. 
 
(4)

 

AUX A connector 

BNC connector for a user defined signal. Standard is not connected. 
 
(5)

 

DC OUT connector 

BNC  connector  providing  the  measured  DC  signal  with  a  fixed  amplification  of  10.  This 
signal  is  not  filtered.  Use  of  this  connector  allows,  for  instance,  measuring  the  signal  AC 
(signal is provided at the front panel) while watching the offset simultaneously.  
 
(6)

 

NOTCH ON A switch 

Switch that enables (upper position) or disables (lower position) the NOTCH filter (optional). 
 

2.5. 

Headstage (Option) 

 

 

 

Figure 3: headstage (optional) and electrode holder (optional) and of the EXT-02F 

Headstage Elements 

 

BNC connector for the electrode holder (measuring electrode) 

REF: SMC connector for the reference electrode 

GND: Ground connector 

holding bar 

 
On  request,  P

EL

  can  be  implemented  using  1  mm  or  2  mm  banana  jacks  or  using  SMC 

connectors. 
 
Very Important: EXT-02F headstages are always labeled “EXT-02” (see Figure 3) and must 
not be exchanged with headstages from other npi electronic EXT amplifiers, e.g. the modular 
EXT-10 2F or EXT-10C which are labeled “EXT”! 
 
Also  Important:  The  shield  of  the  BNC  connector  of  the  headstage  is  connected  to  driven 
shield, and must not be connected to ground. 
 

Summary of Contents for EXT-02F

Page 1: ...E EXT 02F EXTRACELLULAR AMPLIFIER WITH FILTERS One or Two Channel Recordings VERSION 4 2 npi 2015 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 s...

Page 2: ...______ version 4 2 page 2 Table of Contents 1 Safety Regulations 3 2 EXT 02F 4 2 1 Components 4 2 2 System Description 4 2 3 Description of the Front Panel 4 2 4 Description of the Rear Panel 8 2 5 He...

Page 3: ...e check for appropriate line voltage before connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact connected to ground protective earth Befor...

Page 4: ...of the two recording channels are identical The EXT 02F has a differential input with high input impedance to avoid noise The output voltage signal can be HIGH PASS or LOW PASS filtered and is availa...

Page 5: ...osition switch for selecting the corner frequency of the single pole HIGHPASS filter with 6 dB octave In OFF position the HIGHPASS filter is disabled DC recording Corner frequencies Hz 3 30 100 300 40...

Page 6: ...gnals greater than 1 V will be converted into a sound NOISE With an amplification of 10000 this correspondents to a signal of 100 V at the electrode 9 INPUT MODE switch Switch for selecting the INPUT...

Page 7: ...he HIGH PASS FILTER to 400 Hz Set the GAIN range switch to 1k and adjust the baseline to zero using the trimpot Set the GAIN range switch to 10k and adjust the baseline again if necessary 14 INPUT swi...

Page 8: ...are used the headstage must be disconnected Also very Important Setting both switches to OFF means that both inputs are grounded i e measurement is impossible This applies also if a headstage is used...

Page 9: ...ON A switch Switch that enables upper position or disables lower position the NOTCH filter optional 2 5 Headstage Option Figure 3 headstage optional and electrode holder optional and of the EXT 02F He...

Page 10: ...al input that minimizes noise pick up Differential means that the signal for the amplifier is the difference between the positive PEL at the headstage and negative REF at the headstage input of the am...

Page 11: ...______________________________ version 4 2 page 11 Figure 4 headstage connections A differential measurement B single ended measurement Hint The REF connector of the headstage can also be set to GND b...

Page 12: ...Electrical Activity of Individual Neurons In Situ Extra and Intracellular Recording in U Windhorst and H Johansson eds Modern Techniques in Neuroscience Research Springer Berlin New York Lambert F M...

Page 13: ...400 500 600 800 1k 3k 10k with OFF position LOWPASS Filter single pole attenuation 6 dB octave corner frequencies Hz 30 100 300 500 700 900 1 3k 1 5k 1 7k 3k 10k 20k NOTCH Filter Europe 50 Hz USA 60...