Development and manufacturing of non-
destructive testing units and system
Ultrasonic Flaw Detector NOVOTEST UD2301
18
the current point of TVG. The distance
between two neighboring points is not
less than 3 microseconds.
+dВ
Gain of the current point TVG. The total
value of all the points of the TVG not
more 40 dB, accuracy of setting 0.1 dB.
DGS
GAIN
Setting of gain of received path from 0
to 126 dB, in steps of 0.5, 5 dB.
DGS
Turn on / off DGS.
POINT
Select Point TVG and display the total
number of points. The maximum
number of points TVG - 15.
POS.
Displays coordinates of the position of
the current point of TVG. The distance
between two neighboring points is not
less than 3 microseconds.
dВ
Gain of the current point DGS. The total
value of all the points of the DGS not
more 40 dB, accuracy of setting 0.1 dB.
А LEVEL
Allows to set the level / amplitude of the
reference signal.
mm
2
Allows to set the value of the equivalent
area of the reflector.
AFS
START
Coordinate of the beginning of the
selected gate. It can take values from 0
to the maximum value of scanning.
WIDTH
The width of the selected control gate
along the beam. It can take values from
0 to the maximum value of scanning.
The total value of the start and the width
of the gate can not exceed the value of
the maximum scanning.