Operation Manual
Hardness Tester NOVOTEST T-UD3
Page. 32
5.
Moving the cursor with <
> key, select the first line. Make about five measurements on
the first reference sample, the device will show the average value of the series according to the
current calibration (Figure 2.37). If get obvious error of measurement press <
>
key, and
the last measurement will be deleted from the seriers.
Figure 2.37 – Calibration measurements
Press <
>, and using <
> and <
> set the nominal hardness value of the
reference sample.
6.
Use <
> key to move to the next value, press <
> to save the first one. After the
correction of the hardness value of the first sample the display will look like on Figure 2.38.
Figure 2.38 - After correcting the first sample
7.
Press <
>, to set up the second sample value, the procedure will be the same as for the
first one. After the end of all the operations, the display will look as on Figure 2.39.