Battery indicator is
Yellow though the
battery gauge is
showing the battery is
full
1. Battery malfunction
2. Device malfunction
1. Replace battery with an
additional charged battery.
2. Connect the original battery to
the charger.
If problem continues across
multiple batteries OR if one of
batteries would not charge or
causes the charger LED to turn
red call your device support
specialist.
Redness of the skin
under the transducer
arrays
Common side effect
1. Use over-the-counter 0.1%
hydrocortisone cream when
switching transducer arrays.
2. Shift transducer arrays 3/4 of
an inch from the last location (so
the adhesive gel is between the
red marks).
If the redness gets worse: See
your doctor.
Blisters under the
transducer arrays
Rare side effect
See your doctor for a prescription
antibacterial cream. Use as your
doctor tells you.
Itching under the
transducer arrays
Rare side effect
1. Use over-the-counter 0.1%
hydrocortisone cream when
switching transducer arrays.
2. Shift transducer arrays over
3/4 of an inch from the last
location (so the adhesive gel is
between the red marks).
If the itching gets worse: See
your doctor.
Pain under the
transducer arrays
Rare side effect
Stop treatment.
See your doctor.