29
The insulin should fill the bottom part of the outer needle
cap. This indicates the device is functioning properly.
If the insulin does not fill or overfills this part of the cap,
review the function check procedure. Then repeat the
function check with a new NovoFine disposable needle and
outer needle cap.
If the second function check also shows under- or overfilling,
do not use your NovoPen Junior.
DO NOT try to repair a NovoPen Junior that you think is not
working properly.
See Warranty section for further information.
20
FUNCTION CHECK (cont.)
5
Hold the NovoPen Junior so
the NovoFine disposable
needle is pointing down.
6
Slowly press the push button
as far as it will go.
7
Check the dose indicator
window to see if it shows
zero (0). If it does not show
zero (0), there is not enough
insulin in the PenFill cartridge
to do a function check. Insert
a new PenFill cartridge (see
pages 8-10) and repeat the
function check.
If there is enough insulin in
the cartridge but the dose
indicator window does not
show zero, repeat the
function check. If you do
not see zero after repeating
F-3
the above steps, do not use
your NovoPen Junior.
Contact Novo Nordisk Inc.
at our toll free number
(1-800-727-6500).