5.4

Relative Humidity

Relative humidity has an adjustment screw located at the rear of the meter.  Obtain the
local relative humidity by using a mechanical or an electronic psychrometer.  Adjust the
screw to obtain the same reading on the humidity dial.  Make only minor adjustments to the
meter at any one time.  Allow the meter to stabilize for a few minutes and then repeat
should more adjustment be required.  An agreement of three to five percent RH should be
acceptable for this type of meter.

5.5

Barometric Pressure

The barometer should be adjusted for the station elevation.  Obtain the local barometric
pressure by calling the local airport or by listening to local weather forecasts.  Adjust the
screw at the rear of the barometer to obtain the same reading.  Follow the local readings
for several days and readjust the meter as required.  Make only minor adjustments at any
time.    Do  not  use  force  or  turn  the  screw  by  a  large  amount  to  avoid  damaging  the
barometer.

6.0

MAINTENANCE

Maintenance of the weather station is limited to keeping the display panel clean and free
of dust.  Use only a damp cloth and water to clean the surfaces of the display panel.  The
sensors should be free of debris and kept clean.  The wind sensor should be checked for
worn bearings, damaged cable, and loose hardware.  Periodic inspections should be made
to help maintain the accuracy of the data.  Each time a reading is made, check the wind
sensor to ensure that it is rotating freely in the wind and there are no major problems.

6

Summary of Contents for 100-1950

Page 1: ...NOVALYNX CORPORATION MODEL 100 1950 WEATHER STATION INSTRUCTION MANUAL...

Page 2: ...ive items If packing facilities are not available take the equipment to the nearest Parcel Post UPS or freight service and obtain assistance with the packaging Write the RA on the outside of the box W...

Page 3: ...tric Pressure 1 2 0 SPECIFICATIONS 2 3 0 INSTALLATION 2 4 0 THEORY OF OPERATION 3 4 1 Temperature 3 4 2 Barometric Pressure 4 4 3 Relative Humidity 4 4 4 Wind Speed 4 4 5 Wind Direction 4 5 0 CALIBRAT...

Page 4: ...emometer circuit is self generating while the vane circuit requires an external power supply 1 2 Temperature The other meters are mechanical devices The temperature dial is connected to a remotely loc...

Page 5: ...weather station is relatively simple The wind sensor and temperature sensor are mounted outdoors with the wind sensor placed onto a mast and the temperature sensor clamped onto a wall The wind sensor...

Page 6: ...nds Clamp or tie down the wind sensor cable every two to three feet using either plastic wire ties or other appropriate hardware to suit the installation and the supporting structure Secure the temper...

Page 7: ...d is sensed by anemometer cups that turn as the wind speed increases above the starting threshold of the sensor The cups are coupled to an AC generator that generates an AC signal as the cups rotate T...

Page 8: ...nemometer bearings at least annually and especially in areas that experience constant wind activity 5 2 Wind Direction As the vane rotates in a clockwise direction as viewed from above the sensor the...

Page 9: ...rt or by listening to local weather forecasts Adjust the screw at the rear of the barometer to obtain the same reading Follow the local readings for several days and readjust the meter as required Mak...

Page 10: ...THER STATION PART NO DESCRIPTION QTY 100 2130 COMBINATION WIND SENSOR 1 100 19501 TEMPERATURE DIAL ASSEMBLY 1 92080000 BAROMETRIC PRESSURE DIAL 1 92080200 HUMIDITY DIAL 1 92080300 WIND SPEED METER 1 9...

Reviews: