When the test is finished, the unit will not function for
ten seconds. Any attempt to use the functions will
result in a beep.
If the fault “Auto calibration impossible” (Er.02) occurs
during the calibration test, restart the operation.
However, if the pressure exceeds the maximum limit
(Excessive cuff pressure - Er. 09) the test will not be
affected.
We recommend carrying out the calibration test at least
once a year, in order to avoid any loss in performance.
8.5. Replacing the safeguard battery
The lithium safeguard battery is indispensable to the correct
functioning of the Diasys Integra Access. When it is
completely run down the message LITH is displayed (see
chapter 5.2.2). It can only be replaced in an authorised
workshop.
Should this message be displayed, contact your distributor,
who will tell you where to send the unit.
8.6. Storage and dispatching
Remove the battery if the Diasys is to be stored for more
than a few days.
The Diasys Integra Access and the battery charger are
delivered in a protective packaging. Check that the unit is
intact before use (bad transport conditions could damage
unit performance). Keep all packaging should the unit need
to be transported at a later date.
38 •
Diasys Integra Access - Manual - GB - Rev 2
Manuel Access GB.qxd 24/02/2004 11:32 Page 38