32 •
Diasys Integra Access - Manual - GB - Rev 2
6
.
9. Display codes
As well as the display of the functioning mode, any
problems occuring during a measurement are indicated by
the display of an error code (see below) which is also noted
as such on the printed report if the software is not used.
If the DiasySoft software is used, these problems will be
detailed both on the computer screen and on the report.
Display
Meaning
Comment
OSCI
Oscillometric mode
COMM
Communication established with printer or computer
MOD
Communication established with modem
GOOD
Modem transmission successful
TEST
Calibration test
STOP
Manual stopping of monitoring
Stop key held down
Er 01
Defective solenoid valve
Transitory electrical problem. If problem
continues, contact your distributor.
Er 02
Auto Calibration impossible
Residual pressure in cuff. Check air tubing.
If problem continues, contact your distributor.
Er 03
Abnormal inflation
Tubing disconnected, bent, pierced or blocked
or cuff too loose. Check the air tubing and
connector and that the cuff is properly wound
around patient’s arm.
Er 04
Measurement not validated
Diastole < 50mm not confirmed
Er 05
Abnormal pressure variation
Muscular contraction or excessive exercise.
Er 07
Excessive measurement time
Measurement time >120 s.
Er 09
Excessive cuff pressure
Er 10
Excessive noise or artefacts
Interference on signal; measurement
impossible
Er 11
Systolic pressure
≤
50mmHg
Er 12
Diastolic pressure
≤
30mmHg
Er 14
Battery run down
Er 15
Not used
Er 16
Stopped by connection
Recording stopped because cable link
connected to Diasys.
Er 17
Measurement stopped manually
Stop key pressed during measurement
Er 18
Stopped by switching mode selector Mode selector switched from measurement
position to programming position
Er 19
Impossible to process oscillometric shapes
Er 25
Heart rate unstable during measurement
Er 26
Error during modem transmission (cf. chapter 7.2)
Er 27
Connection to number called impossible during transmission (cf. chapter 7.2)
Manuel Access GB.qxd 24/02/2004 11:32 Page 32