7.2. Modem Transmission (optional)
This option allows the automatic transmission of the whole
test in progress to a remote computer, via a modem. To
continue with the recording, restart the unit (cf. chapter 6.1)
This requires:
- an M-series Diasys integra Access (M for “Modem
Transmission”)
- a modem that can call out (must be given to the patient
with the Diasys)
- a Diasys-modem cable
- a Computer with a reception modem
- the number to be called (reception modem) entered in the
DiasySoft communication window (version V4.5 or a
above) during programming.
- the reception computer must be on, and the NovaModem
tele-reception module open and waiting reception
The modem must be plugged into the mains and the
telephone socket and switched on before transmission.
Simply connect the Diasys to the modem using the specific
cable to automatically initiate transmission. “Mod” is
displayed during transmission, replaced by “Good” at the
end of a successful transmission.
If “COMM” is displayed, check that the modem is plugged
in and switched on.
If the messages Er26 or Er27 or “COMM” are displayed:
- disconnect the Diasys from the modem
- turn off and switch back on the modem
- reconnect the Diasys to the modem to trigger a new
transmission
34 •
Diasys Integra Access - Manual - GB - Rev 2
Manuel Access GB.qxd 24/02/2004 11:32 Page 34