5.2.2. Safeguard battery (non-rechargeable lithium)
The Diasys Integra Access has a safeguard battery enabling
program settings and data in memory to be protected if the
main battery fails or is missing, thus providing the user with
maximum security.
The safeguard battery will normally last for 3 to 5 years.
It is not accessible to the user. Its status is systematically
tested every time an NiMH battery is placed in the unit, as
follows:
When the battery is inserted, the Diasys Integra Access
automatically tests the safeguard battery and displays one
of the following messages for
3 seconds
either
indicating that the lithium battery is functioning properly,
or
indicating that the lithium battery is not functioning (see
chapter 8 “replacing the safeguard battery”).
5.3. Programming the Diasys integra Access
(mode selector in Programming position)
This manual only concerns manual programming. Software
programming is fully explained in the DiasySoft manual.
We recommend that a recharged battery be inserted each
time before programming. If the battery is run down, the
Diasys Integra Access will display the message (ER 14), if
it is dead, the unit’s keys will be blocked (safeguard mode).