Note: The electrostatic potential mode depends on two infrared lamps focusing to calibrate the detection distance.
The connection network port is only to be connected
to a programmable logic controller (PLC), for
the controlling or regulation of a machine or
installation.
Note :
(1) The operating voltage of
the external power supply
is DC9-12V. If the external
power supply is used, the
DC9V alkaline battery must
be remove.
(2) When the CAN
communication is used to
issue the command to read
the static voltage value, the
green light of the network
port will flash.
(3) Use the external power
supply and press the power
button of the measurer.
When the measurer is
on, the yellow light of the
network port is always on.
4.2. MODE AND DETECTION RANGE
4.3. NETWORK PORT PIN DEFINITION
Number
Detection mode
Range
Measurement error
1
Electrostatic potential mode
±60000V
±5%
2
Ion balance mode
±200V
Number
Colour
Port de connexion
1
Light orange
RS485-A
2
Orange
RS485-B
3
Light green
CANH
4
Blue
CANH
5
Light blue
CANL
6
Green
CANL
7
Light brown (1)
VCC
8
Brown
GND
9
Metal shield housing
PE
10
Green light at the network entrance (2)
Communication indicator
11
Yellow light at the network entrance (3)
External power indicator