Running Glucose Control Solution
8. If test result is above 600 mg/dL or 33.3 mmol/L for glucose,
the screen displays HI with Glu displayed.
If test result is below 20 mg/dL or 1.1 mmol/L for glucose,
the screen displays LO.
Out-of-range results may be caused by the following:
• An error in performing the control test, retest and follow the
instructions carefully.
• The control solution may have expired or have been con-
taminated. Check the expiration date on the control solution
vial. Control solution is good for only 3 months after opening.
Make sure the control solution vial is closed when not in use.
• Expired test strip - Check the expiration date on the test
strip vial.