20
Caution:
The Control Solution range printed on the test strip vial
label is for control solution only. It is used to test the performance
of the Monitoring System and test strip.
If your control solution test results fall outside the range printed
on the test strip vial label:
• The Ketone test strip vial may not be working properly.
• Contact Nova Technical Support or your local supplier.
Testing a Ketone Quality Control Solution
1. Insert a test strip into the Monitoring System. If Monitoring
System was off, the screen displays all segments for 2 sec-
onds then the blinking blood drop symbol and “Ket” appear.
NOTE:
If the strip is removed before you start the test, the screen
goes blank.
Running Ketone Control Solution
Nova Vet Monitor OG PN52899E EN.indd 20
4/14/15 10:26 AM