23
Out-of-range results may be caused by the following:
• An error in performing the control test, retest and follow the
instructions carefully.
• The control solution may have expired or have been con-
taminated. Check the expiration date on the control solution
vial. Control solution is good for only 3 months after opening.
Make sure the control solution vial is closed when not in use.
• Expired test strip - Check the expiration date on the test
strip pouch.
• The test strip may have been damaged. This can be caused
by extreme temperature or by leaving the test strip vial cap
open. Retest using a new test strip.
• Monitoring System malfunction - the Monitoring System may
not be working properly.
Running Ketone Control Solution