2. Press the left/right
buttons to indicate this
sample is a control. (CTL is shown on the display.)
NOTE:
It is important to select control solution test
so the test result does not appear to be one of your
blood test results.
NOTE:
If a test is not performed within 2 minutes
from the insertion of the test strip, the screen goes
blank. To perform a test, take out then replace the
test strip starting from Step 1.
3. Shake the control solution vial. Discard a drop
before use. Squeeze a drop of control solution
onto a clean, hard, dry surface, i.e., control cap.
Running Glucose Control Solution
Nova Vet Monitor OG PN52899E EN.indd 30
4/14/15 10:26 AM