MD 30
10
MD 30 Operating Instructions, Art. No. 31686, V45/12
7
Operation
7.1
Switching the device on and off
The power switch “I/O” (at the rear) is used to switch the control unit on and off.
The device can be switched off at any time irrespective of any procedure for switching off the device.
7.2
Overview: control elements on the operating panel
A)
Display:
Shows the operating values (See chapter “7.3 Overview: Standard display”).
B)
“Program” key:
Selection of program 1 to 10. There are 10 selectable programs for each motor. (To facilitate the
selection of programs, the number of selectable programs can be limited to the ones you need. Refer to
“7.8 Configuration Menu”, Level 1, Programs/Number of programs).
C)
“Motor 1/2” key:
Switching between the connected motors (the green indicator lights beside the motor sockets
indicate the active motor). Pressing the Motor key longer changes direction of rotation.
D)
“Handpiece” key:
Choice of handpieces or contra angle (can be individually deactivated, refer to
“7.8 Configuration Menu”).
E)
“Enter” key:
Enters the configuration menu by pressing long (refer to “7.8 Configuration Menu”).
F)
“Select” keys:
•
By pressing the left “Select” key
“Select
▲
”
the software version is shown.
•
By pressing both “Select” keys at the same time
“Select
▲
+
▼
” the programs will be reset to factory default
settings.
•
The Select keys serve in the configuration menu to change values and parameters
“
▲
”: value adjustment (upwards)
“
▼
”: value adjustment (downwards)
G)
“Speed” keys:
•
Restrict the maximum speed that can be selected using the pedal.
“
▲
”: increases the maximum speed
“
▼
”: reduces the maximum speed
•
By pressing both “Speed” keys at the same time
“Speed
▲
+
▼
” the handpiece calibration will be started
(refer to “7.4.2 Calibrating handpieces”).
H)
“Torque” keys:
•
Restricts the maximum torque.
“
▲
”: increases the maximum torque
“
▼
”: reduces the maximum torque
•
By pressing both “Torque” keys at the same time
“Torque
▲
+
▼
”
the torque modes AL and AS are
exchanged (see chapter “7.5 Torque limit function, AL mode (Automatic Limiter)” and chapter
“7.6 Torque limit function, AS mode (Automatic Stopper)”).
I)
“Pump” keys:
•
Changing the pump flow rate that can be supplied using the pedal.
“
▲
”: increases the maximum supply quantity
“
▼
”: reduces the maximum supply quantity
•
By pressing both “Pump” keys at the same time
“Pump
▲
+
▼
”
the pump will be put on call, pressing again
will switch it off.
A
F
G
H
I
B
C
D
E